Lens Replacement
Only replace the lenses in dry and low-dust conditions.
When the lens is removed, the degree of protection
IP 54
of the system is no longer guaranteed!
Position the thermal imager on a solid, level plane so that
the lens is pointed towards you. Rotate the lens on the
thermography system counterclockwise as far as it will go
and/or until the two red dots are aligned. Then, remove
the lens to the front.
In order to attach the lens, the red marking spots on lens
and camera must be aligned. In this position, the lens
must be pressed slightly against the camera corpus and
turned clockwise until the red dot of the lens is aligned
with the white dot of the camera.
Operation using the Gigabit Ethernet (GigE)
Interface
Please proceed as follows in order to connect the thermal
imager to a PC/notebook. The accessories described are
contained in the scope of delivery as an option.
The Ethernet cable is used in order to establish a
connection to the Ethernet interface of the PC (RJ45
connector).
The 8-pin LEMO connector (blue mark) is connected
to the left 8-pin LEMO socket (Figure 1-2, blue mark)
on the camera.
Caution
Make sure you establish the connection to
the correct LEMO socket. Plug the LEMO
connector into the socket with the red dot
directed to the top. Do not force the
connector in.