Product Manual

Energy Logger
Environmental Specifications
57
Safety
IEC 61010-1
IEC Mains Input...............................................Overvoltage Category II, Pollution Degree 2
Voltage Terminals ...........................................Overvoltage Category IV, Pollution Degree 2
IEC 61010-2-033.................................................CAT IV 600 V / CAT III 1000 V
Electromagnetic Compatibility (EMC)
International ........................................................IEC 61326-1: Industrial
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency
energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly
connected to a low-voltage power supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments due
to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to
a test object.
Korea (KCC)........................................................Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller
or user should take notice of it. This equipment is intended for use in business environments and not to
be used in homes.
USA (FCC) ..........................................................47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Wireless Radio with Adapter
Frequency Range................................................2412 MHz to 2462 MHz
Output Power ......................................................<100 mW