Specifications
71
Trigger
Screen Update ....................................................... Free Run, On Trigger
Source .................................................................... A, B
Sensitivity A and B
@ DC to 5 MHz ................................................... 0.5 divisions or 5 mV
@ 40 MHz
125B, 124B ..................................................... 1.5 divisions
123B ................................................................ 4 divisions
@ 60 MHz
125B, 124B ..................................................... 4 divisions
123B ................................................................ NA
Slope ....................................................................... Positive, Negative
Advanced Scope Functions
Display Modes
Normal ................................................................. Captures up to 25 ns glitches and displays analog-like persistence waveform.
Smooth ................................................................ Suppresses noise from a waveform.
Envelope ............................................................. Records and displays the minimum and maximum of waveforms over time.
Auto Set (Connect-and-View)
Continuous fully automatic adjustments of amplitude, time base, trigger levels, trigger gap, and hold-off. Manual override by user
adjustment of amplitude, time base, or trigger level.
Dual Input Meter
The accuracy of all measurements is within ±(% of reading + number of counts) from 18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty
+1 %. More than one waveform period must be visible on the screen.