Product Manual

Clamp Meters
Safety Information
5
Connect the common test lead before the live test lead and remove the live test
lead before the common test lead.
Remove all probes, test leads, and accessories before the battery door is
opened.
Keep fingers behind the finger guards on the probes.
Hold the Product behind the tactile barrier.
Replace the batteries when the low battery indicator shows to prevent incorrect
measurements.
Do not use the HOLD function to measure unknown potentials. When HOLD is
turned on, the display does not change when a different potential is measured.
Disconnect power and discharge all high-voltage capacitors before you measure
resistance, continuity, capacitance, or a diode junction.
Remove the input signals before you clean the Product.
Use only specified replacement parts.
When batteries are changed, ensure that the calibration seal in the battery
compartment is not damaged. If damaged, the Product may not be safe to use.
Return the Product to Fluke for replacement of the seal.
Do not use in CAT III or CAT IV environments without the protective cap of test
probe, The protective cap decreases the expose probe metal < 4mm. This
decreases the possibility of arc flash from short circuits.
Do not place magnet inside Category IV panel. Place it outside the panel instead.