Specifications
19
Specifications
Thermal
IR Temperature Range -30 °C to 900 °C (-22 °F to 1652 °F)
IR Accuracy (Calibration
geometry with ambient
temperature 23 °C±2 °C)
0 °C: ± 1 °C or ± 1 % of reading, whichever is greater
(32 °F: ± 2 °F or ± 1 % of reading, whichever is greater)
-10 °C to <0 °C: ±2 °C(14 °F to <32 °F: ±4 °F)
<-10°C: ±3 °C (<14 °F: ±6 °F)
IR Repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater.
Display Resolution 0.1 °C / 0.1 °F
Spectral Response 8 μm to 14 μm
Response Time (95 %) <500 ms
K-Type Thermocouple
Input Temperature Range
-270 °C to 1372 °C (-454 °F to 2501 °F)
K-Type Thermocouple
Input Accuracy (with
ambient temperature
23 °C±2 °C)
<-40 °C: ±(1 °C + 0.2 °/1 °C)
-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
-40 °F: ±1 % or 2 °F, whichever is greater
K-Type Thermocouple
Repeatability
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater
K-Type Thermocouple
Resolution
0.1 °C/0.1 °F
Optical
Distance:Spot
(90 % energy)
60:1(calculated at 90 % energy)
Laser sighting Offset dual laser, output <1 mW
Minimum spot size 19 mm
Operational
Emissivity
Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of
common materials (see Table 3 for list of materials and value)
Probe input KTC (Should display simultaneous probe and IR temperature)
Data storage 99 points
Communication USB 2.0
Operating Altitude 2000 meters
Storage Altitude 12,000 meters
Relative Humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating Temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature -20 °C to 60 °C (-4 °F to 140 °F)
Vibration 2.5 G, IEC 60068-2-6
Weight 0.322 kg (0.7099 lb)
Dimensions 17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W
Power
2 AA Batteries
Battery Life
8 hours with laser and backlight on; 100 hours with laser and backlight
off, at 100 % duty cycle (thermometer continuously on)
Standards and Agency
Approval
EMC: meets IEC61326-1: Portable
Safety Compliance: IEC 60825-1, Class 2
Implement Standard: Q/SXAV 16