User's Manual

700G Series
Users Manual
8
Operation
The subsequent sections tell you how to operate the
Product. Push to turn on the Product.
The analog bar graph at the bottom of the display shows
the applied-pressure level relative to the full range of the
gauge.
Note
If you record a Tare value, the pressure shown
is not the actual pressure applied.
How to Setup the Product
Before you use the product, it is necessary to configure it
for your application. Push to go to the Setup menu.
Each time is pushed, the display goes to the
subsequent function. Push or to change the
parameter value. When a parameter is set, push to
exit the configuration menu or to move to the next
parameter.
Engineering Units
The Product’s default engineering unit shows psi. To
change this, push and to move through the 23
standard engineering units plus one custom unit/scale.
When the necessary unit shows, push or .
Pressure now shows in the chosen engineering units.
See the Specifications section for a list of available
engineering units. See the Supervisory Mode section for
instructions to set up custom units.
Set Auto Off
Auto Off can be set in 1-minute increments from 1 to 30
minutes or you can turn off the function for continuous
Product operation. The Product is configured for 30
minutes. Push and to set the necessary interval.
The “off” position is at the low end of the selections, less
than 1 minute.
Show Battery Voltage
Actual battery voltage and a percent-of-life bargraph
show the battery charge. No adjustments are made in this
parameter.
Display Actual Temperature
The Product is temperature compensated This parameter
shows the temperature measured by the internal sensor.
Push or to show degrees F or C.
Set Damping
Selections are “on” and “off” . Damping smooths
readings from pulsating pressure sources.