User's Manual

RTD/Thermocouple Calibrator
712B Calibration Adjustment
21
Adjust 50 Source
1. Set the range of 8508A 4-Wire measurement to 2 k.
2. Push on the 712B to enter the 50 Source calibration mode.
3. Wait approximately 10 seconds until the 8508A shows a stable reading.
4. Use the arrow keys to input the reading into the 712B.
5. Confirm the input.
hwp02.jpg
Adjust 350 Source
1. Push to enter 350 Source calibration mode.
2. Wait approximately 10 seconds until the 8508A shows a stable reading.
3. Input the reading into the 712B.
4. Confirm the input.
hwp03.jpg