User's Manual

712B/714B
Calibration Manual
22
Adjust 500 Source
1. Set the range of 8508A 4-Wire measurement to 20 k.
2. Push to enter 500 Source calibration mode.
3. Wait approximately 10 seconds until the 8508A shows a stable reading.
4. Input the reading into the 712B.
5. Confirm the input.
hwp04.jpg
Adjust 3500 Source
1. Push to enter 500 Source calibration mode.
2. Wait approximately 10 seconds until the 8508A shows a stable reading.
3. Input the reading into the 712B.
4. Confirm the input.
hwp05.jpg