User's Manual

RTD/Thermocouple Calibrator
712B Calibration Adjustment
27
Adjust 500 Input
1. Push on the 712B to enter 500 input calibration mode.
2. Set the 5522A to output 500 .
Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp12.jpg
Adjust 3500 Input
1. Push on the 712B to enter the 3500 input calibration mode.
2. Set the 5522A to output 3500 .
Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp13.jpg