User's Manual

712B/714B
Calibration Manual
32
Use 8508A to Adjust the 5522A Output
To use the 8508A to calibrate the 5522A output:
1. Connect the voltage output terminals of the 5522A with the voltage input
connectors of the 8508A via the pure copper cables.
2. Set the resolution of DCV measurement mode to 7 1/2-digits or above.
3. Adjust the output of the 5522A until the reading of the 8508A is -10 mV
±0.2 μV, and record the voltage setting of the 5522A (=Voltage
-10mV
Setting).
4. Adjust the output of the 5522A until the reading of the 8508A is 75 mV
±0.2 μV, and record the voltage setting of the 5522A (=Voltage
75mV
Setting).
Note
The two voltage setting records are used to calibrate the UUT.
Adjust mV Source
To calibrate the 714B mV Source function:
1. Connect the 714B TC mini plug and the 8508A with pure copper cables, as
shown in Figure 15.
Fluke 8508A
Fluke 714B
Pure Copper TC Miniplug
Pure Copper Cables
hwp006.eps
Figure 15. Connect the 714B to the 8508A
2. Push to enter -10 mV Source calibration mode.
3. Input the reading on the 8508A to the 714B.
hwp21.jpg