User's Manual

Manual Supplement 712B Users
2 1/15
On page 17, add the following to the RTD Types Accepted table:
RTD Type Ice Point (R0) Material α Range (°C)
Pt100 (3902) 100 Platinum 0.003902 /°C -200 to +500
On page 31, add the following to the RTD Input and Output table:
RTD Type
(α)
Range
(°C)
Measure (°C) Source (°C)
1 year 2 year
Source
Current
1 year
2 year
100
Pt(3902)
-200 to +100 0.2 °C 0.4 °C
1 mA
0.2 °C 0.4 °C
100 to 500 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
Change #3, 487
On page 32, replace Electromagnetic Environment with:
Electromagnetic Environment
IEC 61326-1, Portable
FCC: CFR Title 47, Part 15, Subpart B