Data Sheet
3 Fluke Corporation Fluke 71XB Temperature Calibrators
RTD input and output (Fluke 712B)
RTD Type (α) Range (°C)
Measure (°C) Source (°C)
1-year 2-year
Source
current
1-year 2-year
10Ω
Pt(385)
-200 to 100 °C 1.5 °C 3 °C 1 mA 1.5 °C 3 °C
100 to 800 °C 1.8 °C 3.6 °C 1 mA 1.8 °C 3.6 °C
50Ω
Pt(385)
-200 to 100 °C 0.4 °C 0.7 °C 1 mA 0.4 °C 0.7 °C
100 to 800 °C 0.5 °C 0.8 °C 1 mA 0.5 °C 0.8 °C
100 Ω
Pt(385)
-200 to 100 °C 0.2 °C 0.4 °C
1 mA
0.2 °C 0.4 °C
100 to 800 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
200 Ω
Pt(385)
-200 to 100 °C 0.2 °C 0.4 °C
500 μA
0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
500 Ω
Pt(385)
-200 to 100 °C 0.3 °C 0.6 °C
250 μA
0.3 °C 0.6 °C
100 to 630 °C 0.015 %+0.28 °C 0.03 %+0.56 °C 0.015 %+0.28 °C 0.03 %+0.56 °C
1000 Ω
Pt(385)
-200 to 100 °C 0.2 °C 0.4 °C
250 μA
0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
100 Ω
Pt(3916)
-200 to 100 °C 0.2 °C 0.4 °C
1 mA
0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
100 Ω
Pt(3926)
-200 to 100 °C 0.2 °C 0.4 °C
1 mA
0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
10 Ω
Cu(427)
-100 to 260 °C 1.5 °C 3 °C 1 mA 1.5 °C 3 °C
120 Ω
Ni(672)
-80 to 260 °C 0.15 °C 0.3 °C 1 mA 0.15 °C 0.3 °C
50 Ω
Cu(427)
-180 to 200 °C 0.4 °C 0.7 °C 1 mA 0.4 °C 0.7 °C
100 Ω
Cu(427)
-180 to 200 °C 0.2 °C 0.4 °C 1 mA 0.2 °C 0.4 °C
YSI400 15 to 50 °C 0.2 °C 0.4 °C 250 μA 0.2 °C 0.4 °C
1. Sensor inaccuracies not included.
2. Resolution: 0.1 °C.
3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and
Cu50), 0.4 °C (Other RTD types) to the specifications.
4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω),
excitation current (0.25 mA for Pt1000 range).
5. Temperature Coefficient: ±0.05 °C /°C for measure, ± 0.05 °C /°C (<18 °C or >28 °C) for source.
6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms.