Application Note

1
The Fluke 707Ex is an intrinsically safe loop
calibrator
.
Loop calibration
and maintenance
Introduction
Process instrumentation
requires periodic calibration
and maintenance to ensure that
it is operating correctly. This
application note contains
infor
mation to help g
uide tech-
nicians
through some of the
more common loop calibration
tasks using Fluke tools.
Fluke loop calibrators are
ideal for a wide variety of cali-
bration applications. They
include:
Fluke 707 and 707Ex
1
Loop
Calibrators
Fluke 705 Loop Calibrator
Fluke 715 Volt/mA Calibrator
Fluke 787 and 789
ProcessMeter™ tools
A summary of the features of
Fluke’s loop calibrators and a
glossary of terms are at the end
of this application note.
Field testing loop
powered isolators
The loop-powered isolator and
the two-wire isolating trans-
mitter are two of the more
prevalent dev
ic
es in use in
4-20 mA control loops today
(see Figure 1). The testing and
troubleshooting proc
edures for
each are different and need to
be understood by the techni-
cians performing operational
checks on these units in the
field.
Loop-powered isolators
The main purpose of a loop
isolator is to eliminate ground
loops in control systems while
sending the control signal
current to another part of the
system
. L
oop-powered
isolators, unlike two-wire
transmitters, draw their operat-
ing power from the “input” side
of the isolator (see Figure 1),
which requires a pickup voltage
from 5.5 V to 13.5 V depending
on the manufacturer.
The output of the loop isola-
tor is an electrically isolated
mirror image of the input side
current
. The c
omplianc
e voltage
assoc
iated w
ith the output is
greatly reduced from that of the
input side and ranges around
7.5 V. This produces a total loop
loading capability of 350 .
This limited loop drive capabil
-
ity is the primar
y limitation of
the loop isolator.
Two-wire transmitters
Isolating two-wire transmitters
provide the same isolating
functionality as loop isolators
with the added advantage that
many provide signal condition-
ing for a variety of inputs such
as thermocouple, frequency, dc
current, RTD, strain gauge, and
other process inputs.
The transmitter uses supply
power to convert those input
signals to standard 4 to 20 mA
signals for the transmitter
output to the loop. Typical
power supplies for two-w
ire
transmitters are 24 V or 48 V.
P
ower supplies of this size
allow a sig
nificant loop load
capability on the output.
Output
from
2-wire
transmitter
Loop-powered isolator
Two-wire transmitter
Loop
power
supply
Input types; typical
Thermocouple
Frequency
RTD
DC volts
AC volts
DC current
AC current
Strain gauge
Resistance
Loop power
supply
24 V–48 V
Figure 1
Application Note
From the Fluke Digital Library @ www.fluke.com/library

Summary of content (8 pages)