User's Manual

753/754
Calibration Manual
28
6. Continue through the test points.
7. When you complete the test, set the 5522A to STANDBY.
5522A
CALIBRATOR
Fluke 75X
Fluke 5522A
gso14.eps
Figure 12. Three-Wire RTD Measure Verification Connections
Table 16. RTD Measure Verification (100W Pt (385), Three-Wire Connection)
Input °C (Resistance) 1-Year ( °C ) 2-Year ( °C )
-180 °
(
27.096
)
-179.53 to -180.47 -179.46 to -180.54
100 °
(
138.505
)
99.53 to 100.47 99.46 to 100.54
780 ° (369.712 ) 779.39 to 780.61 779.19 to 780.81