User's Manual

DPC/TRACK
Users Manual
7-6
Interval Value and Interval Units
These fields work in conjunction with the Test Date of the most recent
“Passed” Test Results record corresponding to the Tag ID record so that the
Due Date value is properly updated. The Due Date field will not be
updated unless both of these fields are specified. The Key User or Admin
User ID may edit these fields.
Instrument Type
The choices available within this Pick field are dependent upon the current
selection within the Function field. Furthermore, the choice within the
Instrument Type field controls the appearance of other fields on the
Setup page. For example, if “TC Transmitter” is selected, then among the
available fields on the Setup page are TC Type and RJ Compensation.
If the selection is changed to, say, “RTD Transmitter”, then the TC Type
and RJ Compensation fields disappear. Instead, fields such as RTD Type
and RTD Wires become available. The Key User or Admin User ID may
edit this field.
Problem(s)
When this checkbox is selected, during the test, the system will prompt the
tester to provide additional “human factor” information as to what problems
were observed during the test. You may select none, any, or all of the
checkboxes in this area of the Setup Cal. page.
Mark for Load
This checkbox (YesNo) field may be used in conjunction with your Test
Scheduler and Calibrator Manager tools to help manage your testing needs.
The benefits of using this field are addressed in Chapter 5, “Testing,” and
also within Chapter 9, “System Maintenance.” The Key User or Admin User
ID may edit this field.
Reason for work
When this checkbox (located at the bottom of the Setup Cal. page) is
selected, during the test, the system will prompt the tester to provide
additional “human factor” information as to the reason for work. You may
select none, any, or all of the checkboxes in this area of the Setup Cal.
page.
Tolerance
This Number field is available on the Setup page. The % symbol is part of
the display format for the field; you shouldn’t type this symbol when
changing data in this field. (If you do enter the % symbol, the entered value
will be divided by 100 when you leave the field.)