User's Manual
System Maintenance
Setup Options
9
9-5
Setup Options
Access the Setup Options screen by selecting the System Setup button on the
Main Menu, then selecting the Setup Options... button.
On the General page, you may define the format for Certification information.
Select the Load tab to access the Load page.
The features of this screen are:
Clear marks for loaded Tag IDs
When this checkbox is selected (default setting), DPC/TRACK will
automatically clear the Mark for Load checkbox for Tag IDs just loaded
whose Mark for Load checkbox was selected. Otherwise, the management
of the Mark for Load checkbox is left entirely up to you. (This checkbox
only applies to Tag IDs loaded to the calibrator.)
Always print report (vs. ask user)
When this checkbox is selected, DPC/TRACK will automatically generate a
printed report of the test parameters just loaded to the calibrator. Make sure
your printer is properly attached and turned on if you choose to use this
feature. (This checkbox only applies to Tag IDs as they are loaded to the
calibrator.)
Default Testpoints
The values within this field will be used when the Test Strategy value for a
Tag ID is set to “Defaults”. (The percentage values are relative to the
specified input range. Assuming the device has a 3-15 PSI input range, then
the actual testpoints would be at 3, 9, and 15 PSI if the testpoint values were
0%, 50%, and 100%.) Edit this field per the rules for editing arrays (see
Chapter 7, “Changing Data”).
Select the Unload tab to access the Unload page.
The following checkboxes are included on the Unload page:
Prompt for additional test data after unload
Select this checkbox to have DPC/TRACK prompt for additional post-test
information after an unload is performed.
Set marks for failed Tag IDs
This checkbox is provided as a further means of managing the Mark for
Load checkbox. Selecting the Set marks for failed Tag IDs checkbox
automatically selects the Mark for Load checkbox for any unloaded
“failed” Tag IDs. This way, the next time a load is to be performed, all you
have to do is select the Add Marked button on the Calibrator Manager