screen to insert the Tag ID values for previously “failed” tests into the Load
List. (This checkbox only applies to Tag IDs unloaded from the calibrator.)
Set marks for untested Tag IDs
This checkbox is provided as a further means of managing the Mark for
Load checkbox. Selecting the Set marks for untested Tag IDs
checkbox automatically selects the Mark for Load checkbox for any
unloaded “untested” Tag IDs. The next time a load is to be performed, you
may select the Add Marked button on the Calibrator Manager screen to
insert the previously “untested” Tag IDs into the Load List. (This checkbox
only applies to Tag IDs unloaded from the calibrator.)
Always print report (vs. ask user)
When this checkbox is selected, DPC/TRACK will automatically generate a
printed report of the test results just unloaded from the calibrator, rather than
first prompting the user. Make sure your printer is properly attached and
turned on if you choose to use this feature. (This checkbox only applies to
Tag IDs as they are unloaded from the calibrator.)
Select the Calibration tab to access the Calibration page.
This page allows you to specify your calibration setup options. Besides the
Recalculate Due Date checkbox, this page includes the following area:
TempStd
The radio button selected in this area specifies which definition of
temperature is being used within your process for tests involving
thermocouples or RTDs.