User's Manual
Fluke Calibrators
Performing a Test on the Calibrator B
B-5
Note
This example employs the sample SAMPLETAG6 record that came
with DPC/TRACK. Assuming no data has been changed for this
record, the test should proceed as follows. Specific instructions on
loading the calibrator may be found in Example 2.4. Also refer to
Chapter 5, “Testing.”
1. If necessary, select the Power Key to turn on the calibrator.
2. Repeatedly select Softkey #4 until the word “Tasks” appears above Softkey
#1.
3. Select Softkey #1 to display the list of tasks currently loaded on the
calibrator. One of these tasks should be SAMPLETAG6. (You may have to
select Softkey #3 to scroll to the next page of tasks if SAMPLETAG6 does
not appear on the first page.)
4. If necessary, use the Arrow Keys to position the task selection indicator next
to the SAMPLETAG6 task.
5. Select the Enter Key to choose the SAMPLETAG6 task. Depending upon
your setup options, you may now be presented with a variety of information,
such as Responsible, Procedure Number, etc. Select Softkey #4 until you see
the setup message “This is a Setup Message.”
6. Select Softkey #4 to continue. You should now see the test parameters we
defined earlier for SAMPLETAG6.
7. Select Softkey #4 to continue. The SOURCE portion of the screen should
display the first point to be tested, which is 1 Volt in our case.
8. Attach the two jumper wires to the input and output jacks of the calibrator for
a “Measure Source” test of our wires (as indicated in Figure B-3).