User's Manual

Instrument Record Screen
15.2 Tabbed Notebook Section15
15-9
would like a new point added and select “Edit” & “Insert Before Current Test Point”. To
delete a specific point, click anywhere on that line and select “Edit” & “Delete Current
Test Point”. You will NOT receive a warning. To undo your changes, click on the red X
to cancel your changes.
15.2.2.3 Show DPC Settings
Clicking the Show DPC Settings
button will allow you to configure your Source Mode
and Measure Mode. This provides the compatibility between the signal type and the
Source Mode and Measure Mode
15.2.2.3.1 Source Mode Options
Depending on the Source Mode that you select from the dropdown, different options will
appear: (not all options will be available for all Source Modes)
The following options are available on all SOURCE MODE selections.
Loop Power Source The calibrator supplies loop power at 28 V or 24 V DC
24V
28V NOTE: The 75x Series does not support 28V Loop Power, but attempts
to download devices using 28V will automatically use 24V.
External (Off)
Use DefaultThe setting is controlled by the Options tab of the
Upload/Download screen
Delay (sec) If the process instrument needs more time to settle than the calibrator’s
normal settling time (about 2 seconds) at each new stimulus level, enter that time in
seconds for Delay.
Reading Method
Normal (UEV Off)
UEV On(User Entered Value)
Enables technicians to record calibration results that
were sourced and/or measured by other devices such as panel meters or readout-only
devices.
Use Default – The setting is controlled by the Options tab of the
Upload/Download screen
Unit Type
Custom Units enable you to map one unit to another, such as mV to ºC or ºF.
Allows you to use the Fluke 74x or 75x Series with millivolt output accessories
such as the Fluke 80T-IR Temperature Probe, and to document tests using non-
supported units such as parts per million or revolutions per minute
.
Normal – No additional scaling is performed.
PCT Display – lets you scale the output as a percentage in accordance with the
input requirements.
Damping - smoothing of the last several readings,
On
Off