Data Sheet

General specications
FLUKE-753 Documenting Process Calibrator
FLUKE-754 Documenting Process Calibrator-HART
Standard accessories include: Three sets of stackable
test leads, three sets of TP220 test probes with three
sets of “extended tooth” alligator clips, two sets AC280
hook clips, BP7240 Li-ion battery pack, BC7240 battery
charger, C799 field soft case, USB communication cable,
getting started guide, instruction manual on CDROM,
NIST traceable certificate of calibration, DPC/TRACK 2
sample software that enables upload and printing of
calibration records. Model Fluke-754 includes HART
communication cable. Includes C799 Field Soft Case.
Includes three year warranty.
FLUKE-750SW DPC/TRACK 2 Software
Included with DPC/TRACK software:
Software media, Instruction Manual, USB Cable.
FLUKE-700 Pxx Pressure Modules
Included with each Fluke Pressure Module:
BP-ISO Adapter(s) (except with P29 - P31), Instruction
Sheet, NIST traceable calibration report and data,
one-year warranty.
Accessories
Fluke-700PMP Pressure Pump; 100 psi/7 bar
Fluke-700LTP-1 Low Pressure Test Pump
Fluke-700PTP-1 Pneumatic Test Pump;
400 psi/40 bar
Fluke-700HTP-2 Hydraulic Test Pump;
10,000 psi/700 bar
Fluke-700HTH-1 Hydraulic Test Hose
Fluke-700PRV-1 Pressure Relief Valve Kit for HTP
Fluke-700-IV Current Shunt (for mA/mA
applications)
Fluke-700PCK Pressure Calibration Kit
Fluke-700TC1 TC Mini-Plug Kit, 9 types
Fluke-700TC2 TC Mini-Plug Kit, JKTERS
Fluke-700TLK Process Test lead kit
754HCC Smart instrument
communication cable
BC7240 Battery Charger
BP7240 Li-on Battery Pack
C700 Hard Carrying Case
C781 Soft Carrying Case
C799 Soft Field Case
Ordering information
Data log function
Measure functions: Voltage, current, resistance,
frequency, temperature, pressure
Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings
per minute
Maximum record length: 8000 readings
(7980 for 30 or 60 readings per minute)
Ramp function
Source functions: Voltage, current, resistance,
frequency, temperature
Rate: 4 steps/second
Trip detect: Continuity* or voltage
*Continuity detection not available when sourcing current
Loop power function
Voltage: 26 V
Accuracy: 10 %, 18 V minimum at 22 mA
Maximum current: 25 mA, short-circuit protected
Maximum input voltage: 50 V dc
Note: 250 Ω series resistance is automatically supplied whenever loop power is
enabled on 754.
HART modem interface (754 only)
Maximum input voltage: 30 V dc
Environmental specifications
All calibrator specifications apply from +18 °C to
+28 °C unless stated otherwise.
Operating temperature: -10 °C to 50 °C
Storage temperature: -20 °C to 60 °C
Operating altitude: 3000 m above mean sea level
(9842 ft)
90-day specifications: The standard specification
intervals for the 750 Series are 1 and 2 years.
Typical 90-day measurement and source accuracy can
be estimated by dividing the one year “% of reading
or “% of output” specifications by 2. Floor specifications,
expressed as “% of f.s.” or counts or “ohms” remain
constant.
Ingress protection:
IP-52
Power: Internal battery pack li-ion, 7.2 V, 4400 mAh,
30 Wh; Battery life: Typical usage, >8 hours
Dimensions: 136mm x 245mm x 63 mm
(5.4 in x 9.6 in x 2.5 in)
Weight: 1.2 kg (2.7 lb)
Side port connections:
Pressure module connector
USB connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Safety: Complies with CAN/CSA C22.2 No 1010.1-92,
ANSI/ISA S82.01-1994, UL3111, and EN610-1:1993.
Data storage capacity:
1 week of calibration procedures and results
11 Fluke Corporation 750 Series Brochure
Fluke Corporation
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke. The Most Trusted Tools in the World.
For more information call:
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2004-2013 Fluke Corporation. Specifications subject to change without notice.
Printed in U.S.A. 10/2013 1263323MA_EN
Modification of this document is not permitted without written permission from
Fluke Corporation.