User's Manual

ProcessMeter™
Calibration Adjustment
27
DC Voltage Adjustment
1. Connect the ProcessMeter to the volt/ohm output of the 5500A calibrator.
2. Turn the UUT’s switch to T.
3. Press and hold the Calibration Button for approximately 2 seconds. The unit will
beep (see Figure 11).
Note
Pressing the Calibration Button puts the ProcessMeter into and out of
calibration mode. The ProcessMeter will remain in calibration mode until
the unit is turned off or the calibration button is pressed a second time.
CAL appears in the bottom display when the ProcessMeter is in calibration
mode.
4. Press after each sourced value appears. Do not alter the sourced value while
the display reads
Busy.
Applied voltages:
0 V
4 V
40 V
400 V
1000 V
5. When
Store is displayed, press to store the calibration value.