User's Manual
80 Series V Calibration Manual Supplement
6/15 1
Change #1
On page 16, Table 13, add the following:
F8-
Err
Invalid model. Have Meter serviced.
OPEn Open thermocouple is detected.
Change #2, 39785
On page 8, Table 11, replace the Diode row with the following:
1000 V rms < 7.9 V dc 3.000 V dc 1.0 mA typical
Change #3, 41206, 48958
On page 5, replace Table 2 with the following:
Table 2. Model 87 AC Voltage Function Specifications
Function Range Resolution Accuracy
45 – 65 Hz 30 – 200 Hz 200 – 440 Hz 440 Hz - 1 kHz 1 - 5 kHz 5 - 20 kHz
[1]
[2,4]
600.0 mV 0.1 mV
± (0.7 % + 4)
± (1.0 % + 4)
± (2.0 % + 4) ± (2.0 % + 20)
6.000 V 0.001 V
± (0.7 % + 2)
60.00 V 0.01 V
600.0 V 0.1 V
± (2.0 % + 4)
[3]
unspecified
1000 V 1 V unspecified unspecified
Low pass filter
Same as
45 - 65 Hz
± (1.0 % + 4)
+1 % + 4
-6 % - 4
[5]
unspecified unspecified unspecified
[1] Below 10 % of range, add 12 counts.
[2] The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between
1 and 30 counts. A 30 count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce
a much larger constant error in later measurements.
[3] Frequency range: 1 kHz to 2.5 kHz.
[4] A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
[5] Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.
Change #4, 58804, 62912, 64090, 509, 167
On page 19, Table 14, combine steps 47 and 48 and add footnote 7:
Step Test Function Range 5500A output
Display Reading
83 and 88 87
47
(87 and 88 only)
7
Peak Min/Max
6 V dc 6 Vpp, 2 kHz Sq. Wave,
DC offset 1 V
Max = 3.895 to 4.105 Max = 3.897 to 4.103
Min = -1.897 to -2.103 Min = -1.898 to -2.102
7. The Fluke-83 V does not have a Peak Min/Max function.