User's Manual
Manual Supplement 80 Series V Calibration
6/15 3
Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 55 °C)
Battery Type: 9 V, NEDA 1604 or IEC 6F22
Battery Life: 400 hrs typical with alkaline (with backlight off)
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight: 12.5 oz (355 g)
Weight with Holster and Flex-Stand: 22.0 oz (624 g)
On page 5, under Detailed Specifications, prior to Table 2 add:
All ranges unless otherwise noted: In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts, except
600 μA dc range total accuracy = specified accuracy + 60 counts. Temperature not specified.
Some mobile devices that transmit RF energy may transmit levels that far exceed 3 V/m and may damage sensitive
electronic circuits. To insure the best performance, do not allow a device that is transmitting RF energy in excess of
3 V/m to be within 30 cm of the meter while in use.
On page 27, Table 17, replace the TM1 with the following, and delete both TM3s:
TM1 80 Series V Safety Information (Multi-language) 4271753 1
On page 28, replace Figure 8 with:
MP4
MP2
H2-4 (3)
BT1
MP390-391
MP6
MP22
MP10-11
MP5
Holster
MP15
Tilt Stand
MP8
F2
F1
MP14
S2
MP31
U5
CR6
MP81
MP9
MP41
J1-2
MP27-30
J3
H5-9 (5)
MP13
BT2
Quick
Reference
Guide
Safety
Information
amo015c.eps
Figure 8. 80 Series V Final Assembly