Application Note

2 Fluke Corporation Detecting intermittents with a DMM
To use the MIN MAX AVG
Recording mode:
1. Select the appropriate func-
tion for the desired measure-
ment. (ac volts, dc volts,
resistance, ac current, dc cur-
rent, and frequency)
2. Connect the test leads of the
DMM to the circuit to be
measured. Fluke’s SureGrip
test leads and probes offer a
multitude of circuit attach-
ment methods. Make sure you
do this before activating the
MIN MAX AVG function, other-
wise, the minimum reading
will be whatever the ambient
value is on the unconnected
test leads. This could throw
off your analysis of the
recorded data after the
recording time period expired.
3. Switch to manual range by
pressing the RANGE button. If
necessary, repeat pressing the
RANGE button until the cor-
rect range is indicated in the
DMM’s display. This step is
necessary because the DMM
will automatically switch to
manual range mode when-
ever the DMM is placed in the
MIN MAX AVG recording
mode. Once MIN MAX AVG is
activated, you cannot change
the range setting.
4. Press the
MIN MAX
to activate
the MIN MAX AVG recording
mode. In the Fluke 170 Series,
the display indicates the max-
imum reading and beeps each
time a new low or high value
is detected. In the Fluke 180
Series, the primary display
indicates the maximum read-
ing while the secondary dis-
play indicates the present
measured value.
After insuring the DMM won’t
be disturbed and won’t present
a safety hazard to anyone, you
can leave the DMM unattended
while you focus on other tasks.
At anytime during the recording
period, you can review the
stored readings using the proce-
dure in the next step, or pause
the recording mode without
deleting the stored readings by
pressing the HOLD button. To
continue recording, press the
HOLD button again.
5. View the stored readings by
pressing
MIN MAX
. Each press
of this button causes each of
the stored values (lowest,
highest and average read-
ings) to be sequentially
recalled to the DMM’s display.
Elapse time stamp
The ability to determine when
the lowest and highest reading
was detected could also be use-
ful information in determining
the cause of an intermittent
measurement. The Fluke 180
Series DMMs have the additional
ability of storing the amount of
time that has elapsed between
the start of recording and when
a new minimum or maximum or
average value was stored during
MIN MAX AVG recording. As a
result, each stored minimum,
maximum and average reading
has a “time stamp” associated
with it.
By separately documenting
the time of day you activated the
MIN MAX AVG recording mode,
you can easily calculate the
actual time of day a reading was
detected by the DMM. For exam-
ple, assume you activated the
record mode at 3:07:00 p.m. and
the highest reading displayed
has an elapsed time of 47:05.
You simply add the elapsed time
to the start time to determine
what time of day the highest
reading was recorded. In this
example that would be 3:54:05
p.m. the same day.
Using the MIN MAX AVG
recording mode found in both
the 170 and 180 Series DMMs,
can be useful for some intermit-
tent problems. But it assumes
the circuit point you are con-
nected to will indicate the high-
est or lowest reading when the
problem appears. If the intermit-
tent causes the measured value
to be between the highest and
lowest value, then the MIN MAX
AVG function will not be of
much help in determining the
source of the intermittent.
Advanced MIN MAX
Recording functions
The Fluke 189 True-rms Multi-
meter not only has the stand-
alone MIN MAX AVG Recording
feature described above, but
also incorporates this feature
with another called AutoHOLD
and additional storage memory
to create the Event Logging
function.
AutoHOLD has the ability to
sense when a measured signal
becomes unstable and when it
re-stabilizes. Using the Auto-
HOLD feature to trigger the
starting and stopping of the
MIN MAX record function, you
are not limited to using the DMM
for problems that only result in
signals moving to a minimum or
maximum value.
To use the Event Logging fea-
ture, you will need a Fluke 189
True-rms Multimeter, FlukeView
®
Forms Documenting Software
and a personal computer on
which to download and view
the recorded data. The DMM is
taken to the site where the
recording needs to be done and
set up to record for a period of
time, hopefully during which the
intermittent will occur. Only after
the recording is complete do you
need the software and PC.
As in the MIN MAX AVG
Recording steps above, setup the
meter by selecting the desired
measurement function and set
the range of the meter manually.
Connect the meter’s input to the
point to be measured and then
activate Event Logging.
Tip:
Do not remove the test leads from the cir-
cuit being measured until you have either
pressed the HOLD button to stop the
recording, or looked at and documented
all the stored values. Removing the leads
while recording will result in the DMM
processing the values present on the dis-
connected leads and affect the AVERAGE
value and possibly the lowest or highest
stored values taken during the time the
leads were connected to the circuit.