Data Sheet
Additional Errors
Digits NPLC Additional NPLC Noise Error
C° 0 001 ½ 6
C° 0 01 ½ 6
C° 30.0 1 ½ 5
C° 21.0 2.0 ½ 5
4 ½ 0.02 0.6 °C
Continuity
Continuity Threshold ..................................................Selectable between 1 Ω and 1000 Ω
Test Current...................................................................1 mA
Response Time..............................................................300 samples/sec with audible tone
Accuracy is given as ± (% measurements + % of range)
Range
24 Hour
(23 ±
1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °
C
Outside 18 to 28 °C
1000.0 Ω 0.002 + 0.01 0.008 + 0.02 0.01 + 0.02 0.001 + 0.002
Diode Test
Test Current...................................................................100 μA or 1 mA
Response Time..............................................................300 samples/sec with audible tone.
Accuracy is given as ± (% measurements + % of range)
Range
24 Hour
(23 ±
1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °
C
Outside 18 to 28 °
C
5.0000 V 0.002 + 0.002 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
10.0000 V 0.002 + 0.001 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
Measurement Rates (IEEE488[4])
Measurements/Second
[1]
Function Digits Setting
Integration Time
60 Hz (50 Hz)
8845A 8846A
6½ 100 NPLC1.67 (2) s 0.6 (0.5) 0.6 (0.5)
6½ 10 NPLC 167 (200) ms 6 (5) 6 (5)
5½ 1 NPLC 16.7 (20) ms 60 (50) 60 (50)
5½ 0.2 NPLC 3.3 ms 270 270
DC Volts, DC Current, and
Resistance
4½ 0.02 NPLC 500 us 995 995
6½ 3 Hz 0.47 0.47
6½ 20 Hz 1.64 1.64
AC Voltage and AC Current
[2]
6½ 200 Hz
[3]
4.5 4.5
6½ 1 s 1 1
5½ 100 ms 9.8 9.8
Frequency and Period
4½ 10 ms 80 80
2 AN 2/1 6 ecnaticapaC
[1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
[2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.
[3] For remote operation or external trigger using default settling delay
[4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the
baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus
has a maximum measurement rate of 963 measurement/s.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke. Keeping your world
up and running.
®
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006-2008 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 2/2008 2748044 D-EN-N Rev B
Pub_ID: 11152-eng Rev 02