Data Sheet

8845A and 8846A Fluke Corporation 1
General Specifications
Power
Voltage
100 V Setting............................................................. 90 V to 110 V
120 V Setting............................................................. 108 V to 132 V
220 V Setting............................................................. 198 V to 242 V
240 V Setting............................................................. 216 V to 264 V
Frequency........................................................................ 47 Hz to 440 Hz. Automatically sensed at power-on.
Power Consumption....................................................... 28 VA peak (12 Watt average)
Dimensions
Height .............................................................................. 88 mm (3.46 in.)
Width ............................................................................... 217 mm (8.56 in.)
Depth................................................................................ 297 mm (11.7 in.)
Weight ............................................................................. 3.6 kg (8.0 lb)
Shipping Weight ............................................................ 5.0 kg (11.0 lb)
Display
Vacuum Fluorescent Display, dot matrix
Environment
Temperature
Operating.................................................................... 0 °C to 55 °C
Storage........................................................................ -40 °C to 70 °C
Warm Up..................................................................... 1 hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating.................................................................... 0 °C to 28 °C <90 %
28 °C to 40 °C <80 %
40 °C to 55 °C <50 %
Storage........................................................................ -40 °C to 70 °C <95 %
Altitude
Operating.................................................................... 2,000 Meters
Storage........................................................................ 12,000 Meters
Vibration and Shock......................................................Complies with Mil-T-28800F Type III, Class 5 (Sine only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V
EMC
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has shown
susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz.
Triggering
Samples per Trigger..................................................1 to 50,000
Trigger Delay..............................................................0 s to 3600 s; in 10 µS increments
External Trigger Delay..............................................<1 mS
External Trigger Jitter...............................................<500 µS
Trigger Input..............................................................TTL Levels
Trigger Output............................................................5 V maximum (open collector)
Memory
8845A.........................................................................10,000 measurements, internal only
8846A.........................................................................10,000 measurements, internal, and up to 2 Gigabyte capacity with USB memory module
(available separately, see “Accessories”) through front-panel USB port
Math Functions
Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics (min/max/average/standard deviation), and Limit Test