User's Manual

Front-Panel Operation
Controlling Trigger Functions 3
3-21
4. To store the present Meter configuration in one of the first four memory locations,
press the appropriate soft key. If you want to store the present Meter configuration in
a location other than the first four locations, press the MORE soft key.
The display indicates the next available memory location. If all configuration
memory locations are filled, the Meter will always indicate memory location 10.
5. To adjust the display to the desired memory location, press either <-- or --> to
select a specific digit.
With the digit selected, press the soft key labeled -- to decrement the digit or ++ to
increment the digit.
6. With the preferred memory location set, press the ENTER soft key to store the Meter
configuration.
Storing the Power-up Configuration
To store the Meter’s present configuration as a power-up configuration:
1. Press M.
2. Press the STORE CONFIG softkey.
3. Press the STORE POWER-UP softkey.
The Meter configuration stored as a power-up configuration will be set whenever the
Meter is powered on.
Note
LAN port remote settings (address, hostname, DHCP, mask, etc.) are saved
separately when selected in each instrument, and are not included in
individually stored/copied configurations.
Recalling the Power-up Configuration
In addition to having the power-up configuration to set up the Meter at power on, the
power-up configuration can be recalled through the front-panel keys. To recall the power-
up configuration:
1. Press M.
2. Press the RECALL POWER-UP softkey.
1.4748
F1 F2 F3 F4 F5
RECALL
USB
RECALL
INT MEM
RECALL CONFIG
caw063.eps
3. Press the RECALL CONFIG softkey.
Note
The RECALL POWER-UP softkey will only appear when a power-up
configuration has been saved in the Meter’s memory.