Hart Scientific 9011 High-Accuracy Dual-Well Calibrator User’s Guide Rev.
Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. • Copyright © 2005 • Printed in USA Rev.
Table of Contents 1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 1.2 1.3 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.3.1 1.3.2 1.4 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.1 6.3.2 6.4 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Temperature Display Hold . . . . . . . . . . . . . . . . . . . . . 22 6.4.1 6.4.2 6.5 Switch Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Scan Hold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Ramp and Soak Program Menu . . . . . . . . . . . . . . . . . . .
9 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 45 9.1 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 45 9.1.1 9.1.2 Calibration Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 11 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 49 11.1 11.
Figures and Tables Table1 Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Table 2 Table 2 Table 2 iv International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 2 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Interchangeable Insert Options . . . . . . . . . . . . . . . . . . . . . 16 Controller Operation Flow Chart. . . . . . . . . . . . . . . . . . . . .
1 Before You Start Introduction 1 1.1 Before You Start Introduction The Hart Scientific 9011 High-Accuracy Dual-Well Calibrator may be used as a portable instrument or bench top temperature calibrator for calibrating thermocouple and RTD temperature probes. Calibrations may be done over a range of –30°C to 670°C (–22°F to 1238°F). Temperature display and setability resolution is 0.1 degrees.
9011 Dual-Well Calibrator User’s Guide 1.2 Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
1 Before You Start Safety Information Symbol Description OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances. C-TIC Australian EMC Mark The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark. 1.
9011 Dual-Well Calibrator User’s Guide • DO NOT use this instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the instrument may cause unknown hazards to the user. • Completely unattended operation is not recommended. • Overhead clearance is required. DO NOT place the instrument under a cabinet or other structure. Always leave enough clearance to allow for safe and easy insertion and removal of probes.
1 Before You Start Safety Information grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded three-prong receptacle. The receptacle must be installed in accordance with local codes and ordinances. Consult a qualified electrician. DO NOT use an extension cord or adapter plug. • If supplied with user accessible fuses, always replace the fuse with one of the same rating, voltage, and type.
9011 Dual-Well Calibrator User’s Guide care. Always carry the instrument in an upright position to prevent the probe sleeves from dropping out. The convenient handle allows for hand carrying the instrument. • If a mains supply power fluctuation occurs, immediately turn off the instrument. Power bumps from brown-outs could damage the instrument. Wait until the power has stabilized before re-energizing the instrument. • Allow for probe expansion inside the well as the block heats.
1 Before You Start Authorized Service Centers CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 Telefax: +65 6799-5588 E-mail: antng@singa.fluke.
2 Specifications and Environmental Conditions Specifications 2 2.1 Specifications and Environmental Conditions Specifications Hot Block Cold Block 50°C to 670°C (122°F to 1238°F) –30°C to 140°C (–22°F to 284°F) Accuracy ±0.2°C at 50°C ±0.4°C at 400°C ±0.65°C at 600°C ±0.25°C (insert wells) ±0.65°C (fixed wells) Stability ±0.02°C at 100°C ±0.06°C at 600°C ±0.02°C at –30°C ±0.04°C at 140°C Uniformity ±0.2°C (±0.05°C typical) ±0.05°C (insert wells) ±0.
9011 Dual-Well Calibrator User’s Guide • ambient relative humidity: 15–50% • pressure: 75kPa–106kPa • mains voltage within ±10% of nominal • vibrations in the calibration environment should be minimized • altitudes less than 2000 meters • indoor use only 2.3 Warranty Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment.
3 Quick Start Unpacking 3 3.1 Quick Start Unpacking Unpack the dry-well carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: • 9011 Dry-well • Power Cord • User's Guide with Report of Calibration • 2 inserts • Insert removal tool • RS-232 Cable • 9930 Interface-it Software 3.
9011 Dual-Well Calibrator User’s Guide After using the calibrator, allow the well to cool by setting the temperature to 25°C and waiting 1/2 hour before turning the instrument off. 3.4 Setting the Temperature Section 6.2 explains in detail how to set the temperature set-point on the calibrator using the front panel keys. The procedure is summarized here. (1) Press “SET” twice to access the set-point value. (2) Press “UP” or “DOWN” to change the set-point value.
4 Parts and Controls Rear Panel 4 Parts and Controls The user should become familiar with the dry-well calibrator and its parts (See Figures 1, 2, and 3). 4.1 Rear Panel Power Cord - The removable power cord (not shown), attaches to the back side of the instrument. It plugs into a standard 115 VAC (optional 230 VAC) grounded socket. Figure 1 Back Panel RS-232 - The RS-232 serial port provides a means for connecting the instrument to a computer or a printer using the included serial cable.
9011 Dual-Well Calibrator User’s Guide The area around the calibrator must be kept clear to allow adequate ventilation. The air is directed from the bottom to the top and may be hot. Allow 6 inches of open space around the calibrator to allow adequate ventilation. Fuse Holders - On the rear panel are four user accessible fuse holders. WARNING: Always leave enough clearance to allow for safe and easy installation and removal of probes. 4.
4 Parts and Controls Top Panel 4.3 Top Panel Well Block - The well block openings are located on the top of both the cold and hot sides. These openings are for insertion of interchangeable inserts. The inserts are designed to accept probes of varying diameters. For insert details see Section 4.4, Constant Temperature Block Assembly. Display Hold - One set of Display Hold connectors are available for each block. These connectors allow external switches to be connected and controlled by the instrument. 4.
9011 Dual-Well Calibrator User’s Guide Hot Block 1/16" (1.6 mm) Removable Insert Well 1.45" (37 mm) I.D. Insert “A” 1/2" (12.7 mm) 3/8" (9.5 mm) Cold Block 1/8" (3.2 mm) 3/8" (9.5 mm) 3/8" (9.5 mm) Insert “D” 3 mm Cold block insert shown (Hot block “C” insert includes eight 1/4" wells) 3/16" (4.8 mm) Figure 3 Interchangeable Insert Options 16 4 mm 6 mm 6 mm 4 mm 1/4" (6.35 mm) 3/16" (4.8 mm) 1/4" (6.35 mm) Insert “C” 1/4" (6.35 mm) Removable Insert Well 1.25" (32 mm) I.D. 1/4" (6.
5 General Operation Setting the Temperature 5 5.1 General Operation Setting the Temperature Section 6.2 explains in detail how to set the temperature set-point on the calibrator using the front panel keys. The procedure is summarized here. (1) Press “SET” twice to access the set-point value. (2) Press “UP” or “DOWN” to change the set-point value. (3) Press “SET” to program in the new set-point. (4) Press and hold “EXIT” to return to the temperature display.
6 Controller Operation Well Temperature 6 Controller Operation This chapter discusses in detail how to operate the dry-well temperature controller using the front control panel. Using the front panel key-switches and LED display the user may monitor the well temperature, set the temperature set-point in degrees C or F, monitor the heater output power, adjust the controller proportional band, and program the calibration parameters, operating parameters, and serial interface configuration.
9011 Dual-Well Calibrator User’s Guide Display Temperature SET UP Enable Switch Status DOWN Disable Switch Status Set-Point Resistance SET + DOWN SET + UP EXIT Select Setpoint SET/EXIT Program Menu SET Adjust Setpoint Number of Setpoints SET/EXIT SET Units °C/°F Select Setpoint SET/EXIT SET Select Setpoint EXIT Scan On/Off Secondary Functions SET/EXIT SET + Soak Time EXIT SET Scan Rate Display Power SET/EXIT Program Function Mode SET/EXIT EXIT Set Proportional Band SET/EX
6 Controller Operation Temperature Set-point S Access set-point memory 1 100. Set-point memory 1 location, 100°C currently used To change to another set-point memory press “UP” or “DOWN”. 4 300. New set-point memory 4 location, 300°C Press “SET” to accept the new selection and access the set-point value. Press “EXIT” to continue and to ignore any changes. S 6.2.
9011 Dual-Well Calibrator User’s Guide 6.3 Scan The scan rate can be set and enabled so that when the set-point is changed the dry-well heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the dry-well heats or cools at the maximum possible rate. 6.3.1 Scan Control The scan is controlled with the scan on/off function that appears in the main menu after the temperature scale units.
6 Controller Operation Temperature Display Hold CAUTION: Do not apply any type of voltage to the display hold terminals. Applying a voltage to the terminals may cause damage to the controller. 6.4.1 Switch Status Connect the thermal switch to the display hold terminals on the either the hot or cold side. To enable the Switch Status while displaying the temperature, press "UP" on the hot or cold side.
9011 Dual-Well Calibrator User’s Guide changed states. This feature allows for minimal block temperature scanning times and improved accuracy. Example 2: With the instrument displaying 25°C, set the set-point to 100°C and enable the Scan Hold feature. When the thermal switch changes states at 50°C, the controller automatically sets a new set-point of 50°C and the status flashes either "o or c". Note: There is a delay between the displayed temperature and the actual temperature of the thermal switch.
6 Controller Operation Ramp and Soak Program Menu S 6.5.2 Save new setting Set-points The next parameters are the program set-points. 1 25 First set-point Use the “UP” or “DOWN” buttons to select any of the set-points. 3 10 Third set-point Press “SET” to change the set-point. 10.00 Set-point value Use “UP” and “DOWN” to change the set-point value. 13.50 New set-point value Press “SET” to save the new set-point value. The other set-points can also be set in the same manner.
9011 Dual-Well Calibrator User’s Guide to n) only or both up and down (from set-point n to 1), and also whether the program stops after one cycle or repeat the cycle indefinitely. The table below shows the action of each of the four program mode settings. Function Action 1 up-stop 2 up-down-stop 3 up-repeat 4 up-down-repeat Pf=1 Program mode Use the “UP” or “DOWN” buttons to change the mode. Pf=4 New mode Press “SET” to continue. S 6.5.
6 Controller Operation Secondary Menu instrument must be at temperature and stable prior to taking the set-point resistance reading. To display the Set-point Resistance, press “SET” and “DOWN” simultaneously when the temperature is displayed. When the “SET” and “DOWN” buttons are released the temperature is again displayed. 99.222 6.7 Current set-point resistance setting is displayed Secondary Menu Functions which are used less often are accessed within the secondary menu.
9011 Dual-Well Calibrator User’s Guide 6.9 Proportional Band In a proportional controller such as this, the heater output power is proportional to the well temperature over a limited range of temperatures around the set-point. This range of temperature is called the proportional band. At the bottom of the proportional band, the heater output is 100%. At the top of the proportional band, the heater output is 0 to 100 (hot side) and –100 to 100 (cold side).
6 Controller Operation Controller Configuration Flashes SEC for secondary menu and then displays the heater power SEC 12.0P S Heater power in percent Access proportional band ProP Flashes and then displays the setting Proportional band setting 4.1 To change the proportional band, press “UP” or “DOWN”. 10.0 New proportional band setting To accept the new setting press “SET”. Press “EXIT” to continue without storing the new value. S 6.
9011 Dual-Well Calibrator User’s Guide Flashes HL H=375 and then displays the setting Current High Limit setting Press “UP” or “DOWN” to adjust the setting. H=300 New High Limit setting To accept the new setting, press “SET”. Press “EXIT” to continue without storing the new value. S 6.13 Accept the new High Limit setting Serial Interface Parameters Note: Both sides of the 9011 must be turned on within three seconds of one another for the serial communications to function on the hot side.
6 Controller Operation Serial Interface Parameters The baud rate of the serial communications may be programmed to 300, 600, 1200, 2400 (default), 4800, or 9600 baud. Use “UP” or “DOWN” to change the baud rate value. 4800 b New baud rate Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the next parameter in the menu. 6.13.2 Sample Period (hot and cold sides) The sample period is the next parameter in the menu.
9011 Dual-Well Calibrator User’s Guide Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the next parameter in the menu. 6.13.4 Linefeed (cold side only) The final parameter in the menu is the linefeed mode. This parameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return.
6 Controller Operation Calibration Parameters The calibration parameters are accessed by pressing “SET” after the name of the parameter is displayed. The value of the parameter may be changed using “UP” or “DOWN”. After the desired value is reached, press “SET” to set the parameter to the new value. Pressing “EXIT” causes the parameter to be skipped ignoring any changes that may have been made. 6.14.
9011 Dual-Well Calibrator User’s Guide 38530 Current ALPHA setting To change the ALPHA setting, press “UP” or “DOWN”. 38600 New ALPHA setting To accept the new setting, press “SET”. Press “EXIT” to continue without storing the new value. S 6.14.4 Accept the new ALPHA setting DELTA This probe parameter characterizes the curvature of the resistance-temperature relationship of the sensor. The value of this parameter is set at the factory for best instrument accuracy.
6 Controller Operation Calibration Parameters S Accept the new BETA setting 35
7 Digital Communication Interface Serial Communications 7 Digital Communication Interface This instrument is capable of communicating with and being controlled by other equipment through the digital serial interface. With a digital serial interface, the instrument may be connected to a computer or other equipment. This allows the user to set the set-point temperature, monitor the temperature, and access any of the other controller functions, all using remote communications equipment.
9011 Dual-Well Calibrator User’s Guide down “SET”, then release both buttons to enter the secondary menu. Press “SET” repeatedly until the display reads “PAr”. Press “UP” once, the serial interface menu is indicated with “SErIAL”. Finally press “SET” to enter the serial parameter menu. In the serial parameters menu are the baud rate, the sample period, the duplex mode, and the linefeed parameter. 7.1.2.
7 Digital Communication Interface Interface Commands 7.1.3 Serial Operation The serial communications uses 8 data bits, one stop bit, and no parity. The set-point and other commands may be sent via the serial interface to set the temperature set-point and view or program the various parameters. The interface commands are discussed in Section 7.2. All commands are ASCII character strings terminated with a carriage-return character (CR, ASCII 13). 7.
9011 Dual-Well Calibrator User’s Guide Table 2 Controller Communications Commands Command Format† Command Example† Read current set-point s[etpoint] s Set current set-point to n s[etpoint]=n s=450 Read temperature unit u[nits] u Set temperature units: u[nits]=c/f Command Description Returned Returned Example Acceptable Values Display Temperature set: 9999.99 {C or F} set: 150.
7 Digital Communication Interface Interface Commands Controller Communications Commands continued Command Example† Returned Returned Example Read program set-point soak time pt pt ti: 999 ti: 5 Set program set-point soak time to pt=n n minutes pt=5 Read program control mode pc pc Set program control mode: pc=g[o]/s[top]/c[ont] Command Description Command Format† Acceptable Values 0 to 500 prog: {OFF or ON} prog: OFF GO or STOP or CONT Start program pc=g[o] pc=g Stop program pc=s[top]
9011 Dual-Well Calibrator User’s Guide Controller Communications Commands continued Command Description Set ALPHA calibration parameter to n Command Format† Command Example† al[pha]=n al=0.0038433 Read DELTA calibration parameter de[lta] de Set DELTA calibration parameter to n de[lta]=n de=1.45 Read BETA calibration parameter (cold side only) be[ta] be Set BETA calibration parameter to be[ta]=n n (cold side only) Returned Returned Example Acceptable Values 0.0032 to 0.0042 de: 9.
8 Test Probe Calibration Calibrating a Single Probe 8 Test Probe Calibration For optimum accuracy and stability, allow the calibrator to warm up for 10 minutes after power-up and then allow adequate stabilization time after reaching the set-point temperature. After completing the test probe calibration procedure, allow the well to cool by setting the temperature to 25°C for one-half hour before switching the power off.
9011 Dual-Well Calibrator User’s Guide Speeding up the calibration process can be accomplished by knowing how soon to make the measurement. It is recommended that typical measurements be made at the desired temperatures with the desired test probes to establish these times.
9 Calibration Procedure Calibration Procedure 9 9.1 Calibration Procedure Calibration Procedure Calibration of this instrument should be performed at regularly scheduled intervals by qualified authorized personnel in accordance with your company’s policy. The following is provided as a basis for developing your own procedure for calibrating this instrument. 9.1.
9011 Dual-Well Calibrator User’s Guide 46 3. Set the set-point to 100°C and allow adequate time for the block to reach this temperature and stabilize. Adjust the ALPHA calibration parameter (see Section 6.14.3) to make the block temperature as measured with the standard thermometer match the set-point. The approximate ratio between a change in ALPHA and a change in temperature at 100°C is about 0.00004 to 1. For example, if the block temperature is high by 0.1°C at 100°C, decrease ALPHA by 0.000004. 4.
10 Maintenance 10 Maintenance • This instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment. • If the outside of the instrument becomes soiled, it may be wiped clean with a damp cloth and mild detergent.
11 Troubleshooting 11 Troubleshooting This section contains information on troubleshooting, CE Comments, and a wiring diagram. 11.1 Troubleshooting Problems, Possible Causes, and Solutions In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions. If a problem arises, please read this section carefully and attempt to understand and solve the problem.
9011 Dual-Well Calibrator User’s Guide Problem Possible Causes and Solutions The display shows any error Controller problem. The error messages signify the following problems with the controller. Err 1 - a RAM error Err 2 - a NVRAM error Err 3 - a Structure error Err 4 - an ADC setup error Err 5 - an ADC ready error Err 6 – a defective control sensor Err 7 – a heater error Initialize the system by performing the Factory Reset Sequence describe above.