General Specifications
27
General Specifications
Measured
Parameters
Range
Display
Resolution
Accuracy % of Reading
Temperature
-5 ° to 122 °F
-20 ° to 50 °C
0.1 °F
0.1 °C
±0.9 °C /±1.62 °F from 40 °C to 50 °C
±0.5 °C /±1.00 °F from 5 °C to 40 °C
±1.1 °C /±1.98 °F from -20 °C to 5 °C
Relative
Humidity
10 to 90 % R.H.
non-condensing
0.1 %
±3 % RH from 10 % RH to 90 % RH
Includes 1 % hysteresis
Air Velocity
50 to 3000 fpm
0.25 to 15 m/sec
1 fpm
0.001 m/sec
±4 % or 4 fpm*
±4 % or .02 m/sec* whichever is greater
* Accuracy specification only valid for velocity readings above 50 fpm or 0.25 m/sec.
CO
2
0 to 5000 ppm 1 ppm
Warm up time 1 min (5 minutes for full specification)
2.75 % + 75 ppm
CO 0 – 500 ppm 1 ppm
±5 % or ±3 ppm, whichever is greater,
@ 20 °C and 50 %RH
Additional de-rating over Temperature:
±0.6 %/°C from calibration temperature
±0.6 %/°C < 20 °C
Long term drift < 2 % per month.
Maximum shift in arid or humid storage conditions*: ±0.6 % per day
(Reference storage test conditions:
50 °C, 15 % RH and 30 °C, 95 % RH)
*After the CO calibration seal is removed and the Meter is stored for an extended period in arid or humid conditions, verify that the sensor is within
specification by applying a known concentration of gas using the calibration procedures listed under Calibration.