Datasheet

2 Fluke Corporation Fluke 6200-2/6500-2 Portable Appliance Testers
Test specifications
The accuracy specification for the display range is defined as ±(%reading + digit counts)
at 23 °C ±5 °C, ≤75 % RH. Between 0 °C and 18 °C and between 28 °C and 40 °C,
accuracy specifications may degrade by 0.1 x (accuracy specification) per °C.
The measurement range meets the service operating errors specified in EN61557-1:
1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2.
Power-on test The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency.
Display range: 90 V to 264 V
Accuracy at 50 Hz: ± ( 2% + 3 counts)
Resolution: 0.1 V (1 V - model 6200)
Input impedance > 1 MΩ // 2.2 nF
Maximum input mains voltage: 264 V
Earth bond test (R
pe
) Display range: 0 to 19.99 Ω
Accuracy (after Bond Test zeroing): ± (2.5 % + 4 counts)
Resolution: 0.01 Ω
Test current: 200 mA AC -0 % +40% into 1.99 Ω
25 A AC ± 20 % into 25 mΩ at 230 V
Open circuit voltage: > 4 V AC, < 24 V AC
Bond test zeroing: Can subtract up to 1.99 Ω
Insulation test (R
iso
) Display range: 0 to 299 MΩ
Accuracy: ± (5 % + 2 counts) from 0.1 to 300 MΩ
Resolution: 0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199.9 MΩ)
1 MΩ (200 to 299 MΩ)
Test voltage: 500 V dc –0 % +25 % at 500 kΩ load or (6500-2 only)
250 V dc -0 % +25 % at 250 kΩ load
Test current: > 1 mA at 500 kΩ load, < 15 mA at 0 Ω
Auto discharge time: < 0.5 s for 1 μF
Max. capacitive load: Operational up to 1 μF
Touch current test Display range: 0 to 1.99 mA AC
Accuracy: ± (4 % + 2 counts)
Resolution: 0.01 mA
Internal resistance (via probe): 2 kΩ
Measuring method: Probe
The appliance under test is energized at mains potential.
Substitute leakage
current test
Display range: 0 to 19.99 mA AC
Accuracy: ± (2.5 % + 3 counts)
Resolution: 0.01 mA
Test voltage: 100 V AC ± 20 %
Load/Leakage Test:
Load current
UK
Display range: 0 to 13 A
Accuracy: ± (4 % + 2 counts)
Resolution: 0.1 A
The appliance under test is energized at mains potential.
AU
Display range: 0 A to 10 A
Accuracy: ± (4 % +2 counts)
Resolution: 0.1 A
The appliance under test is energized at mains potential.
Load/Leakage test:
Load power
Display range: 230 V mains
UK 0.0 VA to 3.2 kVA
AU 0.0 VA to 2.4 kVA
Accuracy: ±(5 % + 3 counts)
Resolution: 1 VA (0 to 999 VA), 0.1 kVA (>1.0 kVA)
The appliance under test is energized at mains potential.
Load/Leakage test:
Leakage current
Display range: 0 to 19.99 mA
Accuracy: ± (4 % + 4 counts)
Resolution: 0.01 mA
The appliance under test is energized at mains potential.
PELV test Accuracy at 50 Hz: ± (2 % + 3 counts)
Overload protection: 300 V rms
Warning threshold: 25 V rms