Datasheet
3 Fluke Corporation Fluke 6200-2/6500-2 Portable Appliance Testers
RCD Test: Trip Current
(6500-2 only)
Operational error ±10 %
Nominal 30 mA
Accuracy ±5 %
RCD Test: Trip Time
(6500-2 only)
Standard requirement 61557 Part 6; tolerance of rated test current
0 % to +10 %
Operational error ±10 %
RCD type AC General-Purpose 30 mA
Display range 310 ms
Resolution 0.1 ms
Accuracy 3 ms
Trip time limit at 100 % (30 mA) 300 ms
Trip time limit at 500 % (150 mA) 40 ms
Environmental specifications
Operating temperature 0 to 40 °C
Relative humidity Non condensing < 10 °C
95 % from 10 to 30 °C
75 % from 30 to 40 °C
Safety specifications
Safety rating Complies with EN61010-1 3rd edition
CAT II, 300 V, pol 2
German version only:
DIN VDE0404-1 and DIN VDE0404-2
IEC/EN 61557, part 1, 2, 4, 6, 10
CAT II, 300 V, pol 2
Mechanical and general specifications
Size (LxWxH) 200 mm x 275 mm x 114 mm
Weight 3.13 kg
Power supply 230 V +10 % -15 %, 50 Hz ±2 Hz or
(6500-2 only: 110V +10 % -15 %, 50 Hz ±2 Hz)
Power consumption
(Tester)
13 W typical (idle)
60 W max. during 25A Bond Test
Storage Temperature: -10 °C to 60 °C
Corrosion: 70 °C @ 95 % RH for 5 days max.
Operating altitude 0 up to 2000 m
Sealing IP-40 (enclosure), IP-20 (connectors)
EMC Complies with EN61326-1, Portable
EMI immunity 3 V/m