Datasheet

METER MODE
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The
specified accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F).
Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C.
MAXIMUM RESOLUTION 5,000 counts
VOLTMETER RANGES 500mV, 5V, 50V, 500V, 1,000V
ACCURACY
Vdc ± (0.5 % + 5 counts)
Vac true rms
15 Hz...60 Hz: ± (1 % + 10 counts)
60 Hz...1 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms
dc...60 Hz: ± (1 % + 10 counts)
60 Hz...1 kHz: ± (2.5 % + 15 counts)
OHMS
Ranges 500, 5k, 50k, 500k, 5M, 30M
Accuracy ± (0.6 % + 5 counts)
OTHER METER FUNCTIONS
Continuity Beeper on < 50 30)
Diode test Up to 2.8V
Amps Adc, Aac, Aac+dc using an optional current
clamp or shunt. Scaling factors: 0.1 mV/A ...
100 V/A
Temperature (°C, °F) With optional accessories. Scale factors 1 °C/mV
or 1 °F/mV
Input impedance 1 M ± 1% // 10 pF ± 2 pF
Advanced meter functions Auto/manual ranging, relative measurements
(Zero reference), TrendPlot recording
RECORDER MODE
SCOPE-RECORD- Dual input waveform storage mode.
ROLL MODE
Source and display Input A, Input B, Dual
Memory depth 27,500 points per input.
Each point consist of Min-Max pair.
Min-Max values Min-Max values are measured at high sample
rate ensuring capture and display of glitches.
Recording modes Single sweep, continuous roll, Start-on-Trigger
(through external), Stop-on-Trigger (through
external)
Stop-on-Trigger ScopeRecord mode can be stopped by an
(through External) individual trigger event, or by an interruption of
a repetitive trigger signal.
Horizontal scale Time from start, time of day
Zoom Up to 100x
Memory Up to 2 dual input ScopeRecord waveforms can
be saved for later recall and analysis.
TRENDPLOT Single or dual input electronic paperless
RECORDING chart recorder. Plots, displays and stores meter
and scope measurements.
Source and display Input A, Input B or DMM input
Memory depth 18,000 points record per input. Per record point
a minimum, a maximum and an average
value, plus a date- and timestamp are stored.
Ranges
- normal view 5 s/div to 30 min/div
- in view-all mode 5 min/div to 48 hr/div
(overview of total record)
Recorded timespan Up to 22 days with a resolution of 1
minute
Recording mode Continuous roll for the duration of the
full recordable timespan
Measurement speed 5 measurements per second or more
Horizontal scale Time from start, time of day
Zoom Up to 64x zoom
Memory Up to 2 TrendPlot recordings can be saved for
later recall and analysis.
CURSOR MEASUREMENTS - ALL RECORDER MODES
Source Input A, B or DMM input
Dual vertical lines Min-Max or Average voltage. Time between
cursors
Single vertical line Min-Max or Average voltage. Absolute date and
time or time from start
Time base range 5 ms/div to 1 min/div 2 min/div
Recorded timespan 6 sec to 24 hr 48 hr
Glitch capture 50 ns 250 ns
Sample rate 20 MS/s 4 MS/s
Resolution 200 µsec to 2 sec 4.8 sec
Technical Datasheet Fluke ScopeMeters 3