Datasheet
METER MODE
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The
specified accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F).
Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C.
MAXIMUM RESOLUTION 5,000 counts
VOLTMETER RANGES 500mV, 5V, 50V, 500V, 1,000V
ACCURACY
Vdc ± (0.5 % + 5 counts)
Vac true rms
15 Hz...60 Hz: ± (1 % + 10 counts)
60 Hz...1 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms
dc...60 Hz: ± (1 % + 10 counts)
60 Hz...1 kHz: ± (2.5 % + 15 counts)
OHMS
Ranges 500Ω, 5kΩ, 50kΩ, 500kΩ, 5MΩ, 30MΩ
Accuracy ± (0.6 % + 5 counts)
OTHER METER FUNCTIONS
Continuity Beeper on < 50Ω (± 30Ω)
Diode test Up to 2.8V
Amps Adc, Aac, Aac+dc using an optional current
clamp or shunt. Scaling factors: 0.1 mV/A ...
100 V/A
Temperature (°C, °F) With optional accessories. Scale factors 1 °C/mV
or 1 °F/mV
Input impedance 1 MΩ ± 1% // 10 pF ± 2 pF
Advanced meter functions Auto/manual ranging, relative measurements
(Zero reference), TrendPlot recording
RECORDER MODE
SCOPE-RECORD- Dual input waveform storage mode.
ROLL MODE
Source and display Input A, Input B, Dual
Memory depth 27,500 points per input.
Each point consist of Min-Max pair.
Min-Max values Min-Max values are measured at high sample
rate ensuring capture and display of glitches.
Recording modes Single sweep, continuous roll, Start-on-Trigger
(through external), Stop-on-Trigger (through
external)
Stop-on-Trigger ScopeRecord mode can be stopped by an
(through External) individual trigger event, or by an interruption of
a repetitive trigger signal.
Horizontal scale Time from start, time of day
Zoom Up to 100x
Memory Up to 2 dual input ScopeRecord waveforms can
be saved for later recall and analysis.
TRENDPLOT™ Single or dual input electronic paperless
RECORDING chart recorder. Plots, displays and stores meter
and scope measurements.
Source and display Input A, Input B or DMM input
Memory depth 18,000 points record per input. Per record point
a minimum, a maximum and an average
value, plus a date- and timestamp are stored.
Ranges
- normal view 5 s/div to 30 min/div
- in view-all mode 5 min/div to 48 hr/div
(overview of total record)
Recorded timespan Up to 22 days with a resolution of 1
minute
Recording mode Continuous roll for the duration of the
full recordable timespan
Measurement speed 5 measurements per second or more
Horizontal scale Time from start, time of day
Zoom Up to 64x zoom
Memory Up to 2 TrendPlot recordings can be saved for
later recall and analysis.
CURSOR MEASUREMENTS - ALL RECORDER MODES
Source Input A, B or DMM input
Dual vertical lines Min-Max or Average voltage. Time between
cursors
Single vertical line Min-Max or Average voltage. Absolute date and
time or time from start
Time base range 5 ms/div to 1 min/div 2 min/div
Recorded timespan 6 sec to 24 hr 48 hr
Glitch capture 50 ns 250 ns
Sample rate 20 MS/s 4 MS/s
Resolution 200 µsec to 2 sec 4.8 sec
Technical Datasheet Fluke ScopeMeters 3