Datasheet

RECORDER MODE
TRENDPLOT Dual input electronic paperless chart
RECORDING recorder. Plots and displays the actual, mini-
mum, maximum and average of any measure-
ment.
Source and display Input A, Input A and B
Range 15 s/div till 2 days per division (automatic)
Recorded timespan Up to 16 days with a resolution of 1.5 hours
Recording mode Continuous with automatic vertical scaling and
horizontal time
compression
Measurement speed 2.5 measurements per second
maximum
Horizontal scale Time from start
GENERAL SPECIFICATIONS
CASE
Design Rugged, shock proof with integrated protective
holster
Drip and dust proof IP51 according to IEC529
Shock and Vibration Shock 30g, Vibration 3g according to
MIL-T-28800E, Type III, Class 3, Style B
DISPLAY Bright LCD with CCFL backlight, 35/60 cd/m
2
without/with adapter
Size 72 x 72mm (2.8 x 2.8 inch)
Resolution 240 x 240 pixels
Contrast and brightness User adjustable, temperature
compensated
MEMORY SAVE 20 (Fluke 124) and 10 (Fluke 123)
AND RECALL instrument screens with user set-ups and user
text
REAL-TIME CLOCK Time and date stamp TrendPlot recording
POWER
Line power Country specific line voltage
adapter/battery charger included
Battery power Rechargeable Ni-MH BP130 (installed in Fluke
124) or rechargeable NiCd BP120 (installed in
Fluke 123)
Battery operating time Up to 7 hours using BP130,
up to 5 hours using BP120
Battery charging time 5 hours (Fluke 123),
7 hours (Fluke 124)
Battery power saving Auto power down with adjustable
functions power down time. On screen battery power
indicator
MECHANICAL DATA
Size 50 x 115 x 232 mm
(2 x 4.5 x 9.1 inches)
Weight 1.2 kg (2.5 lb.)
SAFETY
Compliance EN61010-1 (2nd edition) Pollution Degree 2;
UL3111-1;
CAN/CSA C22.2 No. 1010.1;
ANSI/ISA S82.01.
INPUT VOLTAGE RATINGS
Maximum input voltage 600V CAT III
(Maximum voltage between input and reference lead)
Maximum input voltage 600 V CAT III, 1000 V CAT II
using VPS40 Probe
(Maximum voltage between probe tip input and reference lead)
Floating voltage 600V CAT III
Maximum voltage between earth ground and any terminal
(signal input or reference lead)
Maximum voltage between Instrument has common grounds
reference leads connected via selfrecovering fault protection. For
different ground potential measurements
between inputs use DP120 differential voltage
probe.
ENVIRONMENTAL
Operating Temperature 0°C to +50°C
Storage temperature -20°C to +60°C
Humidity 10°C to 30°C, 95% RH non condensing
30°C to 40°C, 75% RH non condensing
40°C to 50°C, 45% RH non condensing
Maximum operating altitude 2,000m (6,500 feet)
3,000m (10,000 feet) voltages 400V
Maximum storage altitude 12 km (40,000 feet)
Electro-Magnetic- Emission EN50081-1
Compatibility (EMC) (EN55022 and EN60555-2)
Immunity EN50082-2
(IEC10004-2, -3, -4, -5)
OPTICALLY ISOLATED PC/PRINTER INTERFACE
To printer Supports HP Laserjet
®
, Deskjet
®
, Epson FX/LQ
and postscript printers via optional PAC91
To PC Transfer instrument settings, screen images and
data, compatible with FlukeView
®
software for
Windows
®
via optional PM9080.
WARRANTY 3 years (parts and labor) on main instrument,
1 year on accessories
Technical Datasheet Fluke ScopeMeters 6