ECG Performance Waveforms
Accuracy
Frequency ....................................................... ±2 %
Amplitude ........................................................ ±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz
Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz
Triangle wave.................................................. 2 Hz
Pulse (63 ms pulse width)............................... 30 BPM and 60 BPM