Datasheet

3 Fluke Corporation Fluke 3000 FC Series Test Tools. The Fluke Wireless Team.
Fluke 3000 FC Series Wireless Multimeter
Detailed specifications (Not compatible with Fluke CNX test tools)
For all specifications: accuracy is specified for one year after calibration, at operating temper-
atures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take
the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
AC voltage
Range
1
Resolution
Accuracy
2,3,4
45Hz to 500Hz 500Hz to 1kHz
600.0mV 0.1mV
1.0% + 3 2.0% + 3
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
1000V 1V
1
All ac voltage ranges are specified from 1% of range to 100% of range.
2
Crest factor of ≤ 3 at full scale up to 500V, decreasing linearly to crest factor < 1.5 at 1000V.
3
For non-sinusoidial waveforms, add ± (2% of reading + 2% full scale) typical, for crest factor up to 3.
4
Do not exceed 10
7
V-Hz.
DC voltage, continuity, resistance, diode test and capacitance
Function Range Resolution Accuracy
mV 600.0mV 0.1mV 0.09% + 2
V 6.000V 0.001V
0.09% + 260.00V 0.01V
600.0V 0.1V
1000V 1V 0.15% + 2
600Ω 1Ω Meter beeps at < 25Ω, beeper detects opens or shorts
of 250μs or longer.
600.0Ω 0.1Ω 0.5% + 2
6.000kΩ 0.001kΩ
0.5% + 1
60.00kΩ 0.01kΩ
60.00kΩ 0.1kΩ
600.0kΩ 0.001MΩ
50.00MΩ 0.01MΩ 1.5% + 3
Diode test 2.000V 0.001V 1% + 2
µF 1000nF 1nF
1.2% + 210.00μF 0.01μF
100.0μF 0.1μF
9999μF
1
1μF 10% typical
1
In the 9999μF range for measurements to 1000μF, the measurement accuracy is 1.2% + 2.
AC and DC current
Function Range
1
Resolution Accuracy
mA AC (45Hz to 1kHZ)
60.00mA 0.01mA
1.5% + 3
400.0mA
3
0.1mA
mA DC
2
60.00mA 0.01mA
0.5% + 3
400.0mA
3
0.1mA
1
All ac current ranges are specified from 5% of range to 100% of range.
2
Input burden voltage (typical): 400mA input 2mV/mA.
3
400.0 mA accuracy specified up to 600mA overload.
Contents
General specifications