Datasheet

4 Fluke Corporation Fluke 3000 FC Series Test Tools. The Fluke Wireless Team.
Frequency
Range Resolution Accuracy
[1]
99.99Hz 0.01Hz
0.1% + 1
999.9Hz 0.1Hz
9.999kHz 0.001kHz
99.99kHz 0.01kHz
1
Frequency is specified up to 99.99kHz in volts and up to 10kHz in amps.
Input characteristics
Function
Overload
Protection
Input Impedance
(nominal) Common Mode Rejection Ratio (1 kΩ unbalance) Normal Mode Rejection
1100 V rms > 10 MΩ < 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
1100 V rms > 10 MΩ < 100 pF > 60 dB, dc to 60 Hz
1100 V rms > 10 MΩ < 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
Open circuit test voltage
Full scale voltage
Typical short circuit current
To 6 MΩ 50 MΩ
1100 V rms < 2.7 V dc < 0.7 V dc < 0.9 V dc < 350 mA
1100 V rms < 2.7 V dc 2.000 V dc < 1.1 mA
Function Overload Protection Overload
mA Fused, 44/100A, 1000V FAST Fuse 600mA overload for 2minutes maximum, 10minutes rest minimum
MIN/MAX recording
Function Accuracy
DC functions The specified accuracy of the measurement function ± 12 counts for changes > 350 ms in duration.
AC functions The specified accuracy of the measurement function ± 40 counts for changes > 900 ms in duration.
General specifications (Not compatible with Fluke CNX test tools)
Maximum voltage between any
terminal and earth ground
1000 V dc or ac rms
Ω fuse protection from A inputs 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
Display (LCD) Update rate: 4/sec
Votls, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts
Battery type Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life 250 hours minimum
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20m
Obstructed, sheetrock wall: Up to 6.5m
Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m
Temperature Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Relative humidity 0 % to 90 % (0 °C to 35 °C), 0 % to 75 % (35°C to 40°C), 0% to 45% (40°C to 50°C)
Altitude Operating: 2,000 m
Storage: 12,000 m
Electromagnetic compatibilty
EMI, RFI, EMC, RF
EN 61326-1:2013, EN 61326-2-2:2013. ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249,
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
Safety compliance ANSI/ISA 61010-1 / (82.02.01): 3
rd
edition
CAN/CSA-C22.2 No 61010-1-12: 3
rd
edition
UL 61010-1: 3
rd
edition
IEC/EN 61010-1:2010
Certifications cCSAus, FCC, CE
Ingress Protection (IP) rating IP54
Pollution Degree 2
Size (H x W x L) 4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in)