Cut Sheet

Fluke 114, 115, 116 and 117
Digital Multimeters
Extended specications
Technical Data
General specications (all models)
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of 0 % to 90 %.
Maximum voltage between any
terminal and earth ground
600 V
Surge protection 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2
W Fuse for A input 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)
Display Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec
Temperature Operating: -10 °C to +50 °C; Storage: -40 °C to +60 °C
Humidity 0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C
Temperature coefficient 0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C)
Operating altitude 2,000 meters
Battery 9 Volt Alkaline, NEDA 1604A/IEC 6LR61
Battery life Alkaline: 400 hours typical, without backlight
Safety compliances ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN
61010-1 2
nd
Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Certifications UL, CSA, TUV, N10140 , VDE
IP rating (dust and water protection) IP42
N10140
Accuracy specications (all models)
Function Range Resolution
Accuracy ± ([% of Reading]
+ [Counts]) Model
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117
DC volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2 114, 115, 116, 117
DC, 45 Hz to
500 Hz
500 Hz to 1 kHz
Auto-V LoZ
1
true-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117
45 Hz to 500 Hz 500 Hz to 1 kHz
AC millivolts
1
true-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
AC volts
1
true-rms 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Continuity
600 W 1 W
Beeper on < 20 W, off > 250 W;
detects opens or shorts of
500 µs or longer
114, 115, 116, 117
Ohms 600.0 W
6.000 kW
60.00 kW
600.0 kW
6.000 MW
40.00 MW
0.1 W
0.001 kW
0.01 kW
0.1 kW
0.001 MW
0.01 MW
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2
114, 115, 116, 117
Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117
Capacitance 1000 nF
10.00 µF
100.0 µF
9999 µF
1 nF
0.01 µF
0.1 µF
1 µF
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF to 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
115, 116, 117
LoZ capacitance (power-up option) 1 nF to 500 µF 10 % + 2 typical 115, 116, 117
1
All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of
range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from
a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3.
AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.