User Manual
1550C/1555
Users Manual
8
Pushbuttons
Use the pushbuttons to control the Tester, view test results, and scroll through
chosen test results. Pushbuttons and their functionality are discussed in
Figure 2.
ENTER
TEST
FUNCTION
1 2 3 4 5 6 7
Ghh02.eps
Item Description
#
Turns the Tester off and on.
$
Push & to go to the Function menu. Push again to exit the
Function menu. To scroll within the Function menu, use the arrow
pushbuttons.
%
Scrolls through test voltages, stored test results, timer duration, and
changes test tag ID characters. Also used to answer “yes” to yes/no
prompts.
&
After a memory location is set, ' displays the test parameters,
test results stored in memory. These include voltage, capacitance,
polarization index, dielectric absorption ratio, and current.
'
Use to scroll through test voltages, stored test results, timer
duration, and memory locations. Also used to answer “no” to yes/no
prompts.
(
Use for Test Voltage mode to start incrementally setting the test
voltage between 250 V and 10,000 V.
)
Starts and stops a test. Push and hold for 1 second to start a test.
Push again to stop a test.
Figure 3. Pushbuttons