User Manual

Insulation Tester
Specifications
31
1000 V < 200 k
200 k to 20 G
20 G to 200 G
> 200 G
unspecified
5 %
20 %
unspecified
2500 V < 200 k
200 k to 50 G
50 G to 500 G
> 500 G
unspecified
5 %
20 %
unspecified
5000 V < 200 k
200 k to 100 G
100 G to 1 T
> 1 T
unspecified
5 %
20 %
unspecified
10,000 V < 200 k
200 k to 200 G
200 G to 2 T
> 2 T
unspecified
5 %
20 %
unspecified
Bar graph range:
Insulation test voltage accuracy:
Induced ac mains current rejection:
Charging rate for capacitive load:
0 to 2 T
-0 %, +10 % at 1 mA load current
2 mA maximum
5 seconds per µF
Discharge rate for capacitive load: 1.5 s/µF
Range Accuracy
Leakage
Current
Measurement
1 nA to 2 mA ±(20 % + 2 nA)
Capacitive
Measurement
0.01 µF to 20.00 µF ±(15 % of reading +
0.03 µF)
Range Resolution
Timer
0 to 99 minutes
Setting: 1 minute
Indication: 1 second
Warning Range Voltage Accuracy
Live circuit
warning
30 V to 1100 V
ac/dc, 50/60 Hz
±(15 % + 2 V)