587 FC/1587/1577 Insulation Multimeters Calibration Manual September 2005, Rev.2, 12/15 © 2005-2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Introduction ............................................................................................ Contacting Fluke ................................................................................... Safety Information ................................................................................. General Specifications .......................................................................... Electrical Specifications .......................................................
1587 FC/1587/1577 Calibration Manual Calibration Adjustment Password...................................................... Changing the Password .................................................................... Restoring the Default Password ........................................................ Meter Keys Used in the Calibration Steps ......................................... Calibration Adjustment ...................................................................... Service and Parts ..............
List of Tables Table Title 1. 2. 3. 4. 5. 6. 7. Symbols..................................................................................................... Required Tools and Equipment ................................................................. DMM Accuracy Tests ................................................................................ Insulation Resistance Accuracy Test ........................................................ Insulation Mode, Source Voltage Accuracy Tests ................
1587 FC/1587/1577 Calibration Manual iv
List of Figures Figure Title 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. Testing the Fuse........................................................................................ Replacing the Batteries and Fuse ............................................................. LCD Display Test ...................................................................................... Disassembling the Meter ........................................................................... Insulation Terminal Clips ..............
1587 FC/1587/1577 Calibration Manual vi
Introduction The Fluke Model 1587 FC, Model 1587, and Model 1577 are battery-powered, true-rms insulation multimeters (hereafter, Meter or UUT) with a 6000-count and a 3 ¾ digit display. Although this manual contains calibration information for Models 1587 FC, 1587, and 1577, all illustrations and examples assume use of the Model 1587 FC. Table 1 provides a description of all the symbols used in this manual. The Meter meets CAT III and CAT IV IEC 61010 standards.
1587 FC/1587/1577 Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. See Table 1 for a list of symbols used on the Meter and in this manual. XW Warning To prevent possible electric shock, fire, or personal injury: 2 • Read all safety information before you use the Product • Examine the case before you use the Product.
Insulation Multimeters Safety Information • Do not apply more than the rated voltage between the terminals or between each terminal and earth ground. • Limit operation to the specified measurement category, voltage, or amperage ratings. • Use Product-approved measurement category (CAT), voltage, and amperage rated accessories (probes, test leads, and adapters) for all measurements. • Measure a known voltage first to make sure that the Product operates correctly.
1587 FC/1587/1577 Calibration Manual Table 1. Symbols Symbol W X T P Ã ) WARNING.RISK OF DANGER. WARNING. HAZARDOUS VOLTAGE. Risk of electric shock. Consult user documentation. AC (Alternating Current) DC (Direct Current) Double Insulated Battery (Low battery when shown on display.) Earth Fuse Conforms to relevant Australian EMC standards. Conforms to European Union directives. Conforms to relevant South Korean EMC Standards. Certified by CSA Group to North American safety standards.
Insulation Multimeters General Specifications General Specifications Maximum Voltage Applied to any Terminal and Common..................................... 1000 V Fuse Protection for mA input .................................. 0.44 A, 1000 V, IR 10 kA Batteries .................................................................. Four AA batteries (NEDA 15A or IEC LR6) Battery Life ..............................................................
1587 FC/1587/1577 Calibration Manual Electrical Specifications Maximum Voltage Applied to any Terminal and Common..................................... 1000 V Fuse Protection for mA input .................................. 0.44 A, 1000 V, IR 10 kA Batteries .................................................................. Four AA batteries (NEDA 15A or IEC LR6) Battery Life ..............................................................
Insulation Multimeters Electrical Specifications Electrical Specifications AC Voltage Measurement (all 1587 models) Range 50 Hz to 60 Hz ±(% of Rdg + Counts) 60 Hz to 5000 Hz ±(% of Rdg + Counts) 600.0 mV 0.1 mV ±(1 % + 3) ±(2 % + 3) 6.000 V 0.001 V ±(1 % + 3) ±(2 % + 3) 60.00 V 0.01 V ±(1 % + 3) ±(2 % + 3) 600.0 V 0.1 V ±(1 % + 3) ±(2 % + 3) [1] ±(2 % + 3) [1] 1000 V [1] Resolution ±(2 % + 3) 1V 1 kHz bandwidth.
1587 FC/1587/1577 Calibration Manual DC Voltage Measurement Range [1] Resolution Specification (1587) [1] ±(% of Rdg + Counts) Specification (1577) [1] ±(% of Rdg + Counts) 6.000 V dc 0.001 V ±(0.09 % + 2) ±(0.2 % + 2) 60.00 V dc 0.01 V ±(0.09 % + 2) ±(0.2 % + 2) 600.0 V dc 0.1 V ±(0.09 % + 2) ±(0.2 % + 2) 1000 V dc 1V ±(0.09 % + 2) ±(0.2 % + 2) Specification applies to both positive and negative polarities. Input Impedance .................................................
Insulation Multimeters Electrical Specifications Continuity Test Continuity Indication ........................................... Continuous audible tone for test resistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω Open Circuit Voltage ........................................... <8.0 V Short Circuit Current ........................................... 1.0 mA typical Overload Protection ............................................ 1000 V rms Response Time .....................................
1587 FC/1587/1577 Calibration Manual Insulation Measurement (1587/1587 FC) Output Voltage Display Range Resolution 50 V 0.01 to 6.00 MΩ 0.01 MΩ (0 % to +20 %) 6.0 to 50.0 MΩ 0.1 MΩ 0.01 to 6.00 MΩ 0.01 MΩ 6.0 to 60.0 MΩ 0.1 MΩ 60 to 100 MΩ 1 MΩ 100 V (0 % to +20 %) 250 V 0.1 to 60.0 MΩ 0.1 MΩ (0 % to +20 %) 60 to 250 MΩ 1 MΩ 500 V 0.1 to 60.0 MΩ 0.1 MΩ (0 % to +20 %) 60 to 500 MΩ 1 MΩ 0.1 to 60.0 MΩ 0.1 MΩ 60 to 600 MΩ 1 MΩ 0.6 to 2.
Insulation Multimeters Basic Maintenance Basic Maintenance This basic maintenance section of the manual contains tests and procedures that require no equipment other than the Meter and some consumables such as fuses and batteries. Also, internal access is limited to the battery and fuse compartment. h Static Awareness h X Semiconductors and integrated circuits can be damaged by electrostatic discharge during handling. This notice explains how to minimize damage to these components. 1.
1587 FC/1587/1577 Calibration Manual Cleaning When cleaning is necessary, wipe the Meter with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture on the terminals can affect readings. Testing the Batteries Under Load XWWarning To avoid electrical shock or personal injury, replace the batteries as soon as the battery indicator () appears. A weak battery can cause false readings. Before testing the batteries, turn the Meter to B.
Insulation Multimeters Basic Maintenance Testing the Fuse XWWarning To avoid electrical shock or injury, remove the test leads and any input signals from the Meter before replacing the fuse. Test the fuse as described below and illustrated in Figure 1. If the fuse is defective, see Replacing the Batteries and/or Fuse later in this manual. 1. Insert a test probe in the J input terminal. 2. Turn the rotary switch to N and verify the Meter is in Auto Range. 3. Insert the probe in the mA input terminal.
1587 FC/1587/1577 Calibration Manual Replacing the Batteries and/or Fuse Use the following procedure and Figure 2 to replace the batteries and/or the fuse: XWWarning To avoid electrical shock, personal injury, or damage to the Meter observe the following warnings: • Replace the batteries as soon as the battery indicator () appears. A weak battery can cause false reading. • Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings specified.
Insulation Multimeters Basic Maintenance Testing the Display The display is an LCD comprised of a series of characters and segments for providing the user with a variety of information. To test the LCD and all of its segments proceed as follows: 1. Set the rotary switch to OFF. 2. Depress and turn the function switch to B. 3. While continuing to depress , verify that the lit display segments match those shown in Figure 3. bav01f.eps Figure 3.
1587 FC/1587/1577 Calibration Manual Disassembling and Reassembling the Meter This section of the manual provides instructions for disassembling and reassembling the Meter. The instructions are limited to major replaceable assemblies and do not include component-level detail. See Figure 4 for an exploded view of the major assemblies. Also, the emphasis is placed on disassembly. When necessary, the procedure includes critical hints for reassembly.
Insulation Multimeters Disassembling and Reassembling the Meter When installing the boot, position the bottom of the Meter in the bottom of the boot, and press the top of the instrument firmly into of the boot. Removing the Battery Door XWWarning To avoid the risk of electrical shock, turn the rotary switch to OFF, and remove the test leads from the front-panel terminals before removing the battery cover. With the boot removed, the next step in disassembling the Meter is to remove the battery door.
1587 FC/1587/1577 Calibration Manual connections by carefully pulling each of the springs back and away from the terminals. See Figure 5. 3. With one hand over the PCA, roll the top case over (face up) and lift it away from the PCA. Note Three red and one black plastic shields are used to isolate the user from the input terminals. With the PCA removed, these shields are loose and can fall away from the PCA. 4. Remove and set aside all four shields for use during reassembly.
Insulation Multimeters Disassembling and Reassembling the Meter Removing the LCD With the PCA removed, the final step in disassembling the Meter is to remove the LCD assembly from the PCA. Refer to Figure 6 and use the following procedure to remove the LCD assembly: 1. Remove the two screws from the display end of the PCA. 2. Hold the PCA face down with the fuse-end of the PCA toward you. Locate the screw-hole near the center of the assembly.
1587 FC/1587/1577 Calibration Manual panel. After the LCD assembly is in position, lock it in place by pressing (below the display) the bezel against the PCA; listen for the tab on the bezel to snap (lock) into position on the PCA. 8. Secure the PCA to the LCD assembly by installing the two screws that attach the PCA to the bezel. These screws also ensure contact between the glass LCD display, the elastomeric strip, and the PCA.
Insulation Multimeters Required Tools and Equipment Required Tools and Equipment Table 2 lists the require equipment for performing the Performance Test and Calibration adjustments. If a recommended model is not available, use a substitute with equivalent or better specifications. Table 2. Required Tools and Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range: 0 V to 1000 V Accuracy: ±0.
1587 FC/1587/1577 Calibration Manual Performance Test The following series of tests comprise a performance test for verifying the accuracy of the Meter (UUT) and its performance level. The performance test is recommended as an acceptance test for incoming inspection and as a calibration procedure for periodically ensuring the accuracy of the Meter. Fluke recommends running the performance test at least once a year.
Insulation Multimeters Performance Test Table 3. DMM Accuracy Tests (cont) UUT Function Step UUT Range UUT Input Level UUT Display Reading 1577 n/a 1587 FC/1587 12 Hz, V ac 99.99 kHz *90 kHz, 2V 13 V dc 6.000 V 0.3 V 0.298 to 0.302 0.298 to 0.302 14 V dc 6.000 V 6V 5.986 to 6.014 5.993 to 6.007 15 V dc 60.00 V 60 V 59.86 to 60.14 59.93 to 60.07 16 V dc 600.0 V -600.0 V -601.4 to -598.6 -600.3 to -599.7 17 V dc 1000 V 1000 V 996 to 1004 997 to 1003 18 mV dc 600.
1587 FC/1587/1577 Calibration Manual Testing Temperature Function (1587 FC/1587 Only) Use the following procedure to verify that the UUT measures temperature within the published specification. 1. Connect the K-Type thermocouple to the temperature input of the UUT (J) and temperature calibrator. Note To ensure an accurate measurement the UUT and thermocouple connector must be at the same temperature. 2.
Insulation Multimeters Performance Test Insulation Resistance Accuracy Test To test the Insulation Resistance accuracy, sequentially complete the test steps in Table 4, using the following procedure. 1. Connect the INSULATION output terminals on the UUT to the calibrator. 2. Set the calibrator to the Insulation Resistance Test Mode. 3. Turn the rotary switch to INSULATION. 4. Set the voltage range on the UUT; see Table 4. 5. Apply the Calibrator Output to the UUT; see Table 4. 6. Press on the UUT. 7.
1587 FC/1587/1577 Calibration Manual Insulation Function, Source Voltage Accuracy Test The following test verifies the accuracy of the source voltage for the insulation function. WCaution To prevent damage to the Test DMM use a HV Probe (Divider) with the Test DMM when testing the 1000V range of the UUT. 1. Connect the output of the HV Probe to the V – Com inputs on the Test DMM and turn the rotary switch to E. 2.
Insulation Multimeters Performance Test I Nominal Test The following test verifies the ability of the UUT to maintain the nominal insulation test current while loaded. 1. Connect the Test DMM and Calibrator to the UUT INSULATION terminals as shown in Figure 8. 2. Set the Calibrator to the Insulation Resistance Test Mode. 3. Set the output resistance on the Calibrator 1 MΩ. 4. Set the Test DMM to measure μA DC 5. Set the UUT to INSULATION, 1000V range. 6. Press on the UUT.
1587 FC/1587/1577 Calibration Manual 7. Turn the rotary switch to OFF to reset the display. UUT (Rear) UUT (Front) 6.0 Vdc Power Supply Power Supply A ebp08f.eps Figure 9. Low-Battery Test Connections Low-Battery Test - INSULATION Function The following procedure verifies the accuracy of the low-battery indicator () during insulation operation. XWWarning To avoid shock hazard DO NOT press during the following procedure. 1. Remove the batteries from the battery compartment. 2.
Insulation Multimeters Calibration Adjustment Procedure Calibration Adjustment Procedure The following sections comprise the Calibration Adjustment Procedure. The procedure is meant to bring the UUT back into specification following repair of the UUT or when the UUT fails the Performance Test. The required equipment is listed earlier in Table 2. The Meter features closed-case calibration adjustment using known reference sources.
1587 FC/1587/1577 Calibration Manual 6. Press the four keys representing the new password. 7. Press to store the new password. Use these steps to change the password on a UUT with SN <32980000: 8. While holding down H on the UUT, turn the rotary switch from OFF to B. The UUT displays Z . 9. Press once to see the calibration counter. 10. Press again to start the password entry. The UUT displays ????.
Insulation Multimeters Calibration Adjustment Procedure SN> 32980000 C51 CR8 T1 R84 C45 R90 R18 R43 R16 C54 R107 R106 C58 CR10 C64 R54 U17 U11 R25 U18 U14 R15 U6 L3 C12 C38 C16 F2 XBT1 R14 Q5 R99 C18 C33 Q7 R51 Q20 C59 R26 R55 C26 C25 U3 U8 R56 R81 Battery Pads CR3 Q21 R36 R76 Q17 Q16 C65 C47 Q22 R80 R65 R63 C71 R27 C72 R64 C20 C14 R34 C29 C31 R117 U7 Q25 R33 C52 U19 C32 R35 C34 C30 U2 C46 R94 R96 R88 S2 R62 CR7 CR9 R97 C53 R91 C37 C39 C40 C41 R40 Y1
1587 FC/1587/1577 Calibration Manual Meter Keys Used in the Calibration Steps The Meter keys behave as follows when performing the Calibration Adjustment Procedure. This may be of help determining why a calibration step is not accepted and for determining the input value without referring to Table 6. H Press and hold to show the measured value. The measurement value is not calibrated so it may not match the input value. This is normal. Press and hold to display the required input amplitude.
Insulation Multimeters Calibration Adjustment Procedure Note Set the calibrator to Standby prior to changing the function switch position and/or after completing adjustment of each function. If the calibration adjustment procedure is not completed correctly, the UUT will not operate correctly. Table 6. Calibration Adjustment Steps for Models with SN <32980000 Switch Position (Function) Input Terminal AC Volts V/COM Calibration Adjustment Step 600.0 mV, 60 Hz (Model 1587 only) [1] 600.
1587 FC/1587/1577 Calibration Manual Table 7. Calibration Adjustment Steps for 1587 FC/1587 Models with SN >32980000 Switch Position (Function) Input Terminal AC Volts V/COM DC Volts Calibration Adjustment Step Input Value 600.0 mV, 60 Hz 600.0 mV, 5 kHz 6.000 V, 60 Hz 6.000 V, 5 kHz 60.00 V, 60 Hz 60.0 V, 5 kHz 600.0 V, 60 Hz 600.0 V, 2 kHz 6.000 V, 0 Hz 60.00 V, 0 Hz 600.0 V, 0 Hz DC Millivolts 600.
Insulation Multimeters Calibration Adjustment Procedure Table 8. Calibration Adjustment Steps for 1577 Models with SN >32980000 Switch Position (Function) Input Terminal AC Volts V/COM DC Volts DC Millivolts Ohms Continuity Diode Test milliamps Insulation mA/COM mA/COM Calibration Adjustment Step Input Value 600.0 mV, 60 Hz 6.000 V, 60 Hz 60.00 V, 60 Hz 600.0 V, 60 Hz 6.000 V, 0 Hz 60.00 V, 0 Hz 600.0 V, 0 Hz 600.0 mV, 0 Hz 60.
1587 FC/1587/1577 Calibration Manual Service and Parts User service is limited to replacing parts. Table 9 identifies the parts available for replacement and Figure 11 shows the location of each part. To order replacement parts refer to Contacting Fluke earlier in this manual. ebp09f.eps Figure 11.
Insulation Multimeters Service and Parts Table 9.
1587 FC/1587/1577 Calibration Manual Table 9. 1587 FC/1587/1577 Replacement Parts (Cont.