All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor ≤ 1.5 at 1000 V.
For non-sinusoidal waveforms, add -(2 % reading + 2 % full scale) typical, for crest factors up to 3.
In the 9999 F range for measurements to 1000 F, the measurement accuracy is 1.2 % + 2 for all models.
Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
400.0 mA accuracy specified up to 600 mA overload.
> 10A unspecified.
Function
Range ±([ % of read-
ing ] + [ counts ])1
Resolution Accuracy
Model 175 Model 177 Model 179
AC volts
600.0 mV 0.1 mV
1.0 % + 3
(45 Hz to 500 Hz)
1.0 % + 3
(45 Hz to 500 Hz)
1.0 % + 3
(45 Hz to 500 Hz)
6.000 V 0.001 V
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V
2.0 % + 3 (500 Hz to
1 kHz)
2.0 % + 3 (500 Hz to
1 kHz)
2.0 % + 3 (500 Hz
to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2
DC volts
6.000 V 0.001 V 0.15 % + 2 0.09 % + 2 0.09 % + 2
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V 0.15 % + 2 0.15 % + 2 0.15 % + 2
Continuity 600 1
Meter beeps at < 25 , beeper turns off at > 250 ; detects opens
or shorts of 250 s or longer.
Ohms
600.0 0.1 0.9 % + 2 0.9 % + 2 0.9 % + 2
6.000 k 0.001 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
60.00 k 0.01 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
600.0 k 0.1 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
6.000 M 0.001 M 0.9 % + 1 0.9 % + 1 0.9 % + 1
50.00 M 0.01 M 1.5 % + 3 1.5 % + 3 1.5 % + 3
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
1000 nF 1 nF 1.2 % + 2 1.2 % + 2 1.2 % + 2
10.00 F 0.01 F 1.2 % + 2 1.2 % + 2 1.2 % + 2
100.0 F 0.1 F 1.2 % + 2 1.2 % + 2 1.2 % + 2
9999 F 1 F 10 % typical 10 % typical 10 % typical
AC amps
(True-rms)
(45 Hz to 1kHz)
60.00 mA 0.01 mA
1.5 % + 3 1.5 % + 3 1.5 % + 3
400.0 mA 0.1 mA
6.000 A 0.001 A
10.00 A 0.01 A
DC amps
60.00 mA 0.01 mA
1.0 % + 3 1.0 % + 3 1.0 % + 3
400.0 mA 0.1 mA
6.000 A 0.001 A
10.00 A 0.01 A
Hz 99.99 Hz 0.01 Hz 0.1 % + 1 0.1 % + 1 0.1 % + 1
(AC- or DC-coupled,
V or A’ input )
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
Temperature
-40 °C to +400 °C
-40 °F to +752 °F
0.1 °C
0.1 °F
NA NA
1 % + 10
1 % + 18
Min-max avg
For DC functions, accuracy is the specified accuracy of the measurement function ± 12 counts for changes longer
than 350 ms in duration.
For AC functions, accuracy is the specified accuracy of the measurement function ± 40 counts for changes longer
than 900 ms in duration
Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 %
to 90 %. Accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] )