Capacitance Measurements
WCaution
To prevent damage to the Product or to the
equipment under test, disconnect circuit
power and discharge all high-voltage
capacitors before you measure capacitance.
Use the dc voltage function to make sure that
the capacitor is discharged.
The capacitance ranges of the Product are 10.00 nF,
100.0 nF, 1.000 μF, 10.00 μF, 100.0 μF, and 9999 μF.
To measure capacitance, set up the Product as shown in
Figure 6.
For the best capacitance measurement accuracy on
capacitance less than 1000 nF, use the relative (REL)
mode to subtract the remaining capacitance of the
Product and leads.
Note
When a capacitor under test has too much
electrical charge, the display shows “diSC“.
+
+
+
+
+
+
+
+
+
Select
Capacitance
grt05.eps
Figure 6. Capacitance Measurements