Ex Safety Information
3
The Product must be completely and
securely fitted in the red holster while it
is in an Ex-hazardous area.
Use only approved accessories with this
Product in Ex-hazardous areas.
Do not use the Product in aggressive
acidic or alkaline solutions. For
applications requiring Group I
equipment, avoid permanent contact of
the Product with oil, hydraulic fluid, or
grease.
Do not use the Product in zones 0, 20,
21, or 22. Measurements on intrinsically
safe connections that go into these
zones are permitted, if the connection
values are met.
XW Warning
To prevent personal injury in mining
hazardous areas:
Avoid extreme mechanical burdens. The
Product can withstand impacts with an
energy of seven joules at -20 °C.
Do not allow the Product to come in
permanent contact with oil, hydraulic
fluid, or grease.
Do not install the Product in a fixed
installation.
XW Warning
To prevent possible electrical shock, fire, or
personal injury in ALL areas of operation:
Read all Safety Information before you
use the Product.
Comply with local and national safety
codes. Use personal protective
equipment (approved rubber gloves, face
protection, and flame-resistant clothes)
to prevent shock and arc blast injury
where hazardous live conductors are
exposed
See the Ex Safety Information section for
additional warnings on Product used in
hazardous areas.
Use the Product only as specified, or the
protection supplied by the Product can
be compromised.