Cut Sheet

Input Characteristics Ranges Accuracy
Input impedance 1 M, 20 pF
Voltage rating 600 Vrms, CAT III
Volt / Amps / Hertz
True-rms voltage (AC+DC) 5.000 V, 50.00 V, 500.0 V, 1250 V* ± (1 % + 10 counts)
True-rms current (AC+DC) 50.00 A, 500.0 A, 5.000 kA, 50.00 kA, 1250 kA ± (1 % + 10 counts)
Frequency 10.0 Hz to 15.0 kHz ± (0.5 % +2 counts)
CF Crest Factor 1.0 to 10.0 ± (5% + 1 count)
Power
W, VA, VAR Reactive Power 250 W 2.50 kW, 25.0 kW, 250kW, 2.50 MW, ± (2 % + 6 counts) Total Power
1-phase and 3-phase, 3 25 MW, 250 MW, 625 MW, 1.56 GW ± (4 % + 4 counts) Fundamental
conductor balanced loads Power
PF Power Factor 0.00 to 1.00 ± 0.04
DPF Displacement Power Factor 0.00 to 0.25 not specified
0.25 to 0.90 ± 0.04
0.90 to 1.00 ± 0.03
Hz Frequency fundamental 40.0 to 70.0 Hz ± (0.5 % + 2 counts)
Harmonics
Volts, Amps, Watts Fundamental V,A ± (3 % + 2 counts),
W ± (5 % + 2 counts)
2 to 31st Harmonic V,A ± (5 % + 3 counts),
W ± (10 % + 10 counts)
32 to 51st Harmonic V,A ± (15 % + 5 counts),
W ± (30 % + 5 counts)
Frequency of fundamental 40 Hz to 70 Hz ± 0.25 Hz
Phase Volt & Amps (between Fund. & Harmonic) 2nd (± 3°) 51st (±15°)
Watts (between Volt Fund. & Amps
Harmonic ) Fund (± 5°) 51st (±15°)
K-Factor (Amps & Watts) 1.0 to 30.0 ±10 %
THD 0.00 to 99.99 ± (3% + 8 counts)
Sags & Swells
Recording times (selectable) 4 min to 16 days
Vrms actual, Vrms max, 5.000 V, 50.00 V 500.0 V, 1250 V* Readings ±(2% +10 counts)
min (AC + DC) Cursor readings ± (2% + 12 counts)
Cursor Readings Average ±(2%
+10 counts)
Arms actual, Arms max, 50.00 A, 500.0 A, 5.000 kA, 50,00 kA
min (AC + DC)
Recording
Recording times (selectable) 4 min to 16 days
Parameters Choose one or two parameters from one of the groups below
V/A/Hz Line Voltage, Current, Frequency
Power Watts, VA, VAR, PF, DPF, Frequency
Harmonics THD, Volts (Fund. & Harmonic), Amps(F&H) Watts(F&H) Freq.(H), %(H) of total, Phase(H), KF
Ohms Ohms, Diode, Continuity, Capacitance
Temperature °C or °F
Scope DC Voltage, DC Current, AC Voltage, AC Current, Frequency, Pulse Width + or -,
Phase, Duty cycle + or -, Peak max, Peak min, Peak min-max, Crest Factor
Transients
Minimum pulse width 40 ns
Useful bandwidth input 1 DC to 1 MHz (with test leads TL24)
Number of transients 40
Voltage threshold settings 20%, 50%, 100%, 200% above or below reference signal
Reference signal After START, the Vrms and frequency of the signal are measured. From these
data a pure sinewave is calculated as reference for threshold setting.
Vpeak min, Vpeak max at cursor 10 V, 25 V, 50 V, 125 V, 250 V, 500 V, 1250 V ± 5% of full scale
*Rated 600V CAT III
Specifications
Accuracies are stated as ± (percentage of
reading + counts) without probes unless
otherwise noted.
Specifications are valid for signals with a
fundamental between 40 and 70 Hz.
Continuously measure volts and amps on a
cycle-by-cycle basis for up to 24 hours
Use cursors to read time and date of sags
and swells
Catch voltage transients and waveform
distortion
Catch and save up to 40 transients
Correlate the cause of transients with time
and date stamps