Datasheet

56x Getting Started Supplement
2/13 1
Change #1, 58989, 64912
On page 10, replace the Specification Summary table
with:
Specifications Summary
See Users Manual on CD for full specifications.
Feature 561 566 568
IR Temperature
Range
-40 C to
550 C
(-40 F to
1022 F)
-40 C to
650 C
(-40 F to
1202 F)
-40 C to 800 C
(-40 F to
1472 F)
Accuracy
<0 C: (1 C + 0.1 /1 C);
0 C: 1 % of reading or 1 C, whichever is
greater
<32 °F: (2 °F +0.1 °/1 °F);
32 °F: 1 % of reading or 2 °F, whichever is
greater
K-Type Thermo-
couple Input
Temperature Range
-40 C to
550 C (-40 F
to 1022 F)
-270 C to 1372 C (-454 F to
2501 F)
K-Type Thermo-
couple Input
Temperature
Accuracy
Input accuracy
1 % of reading
or 2 C (1
% of reading or
4 F),
whichever is
greater
-270 C to -40 C: (1 °C +
0.2 °/1 °C)
(-454 F to -40 F: (2 °F + 0.2
°/1 °F))
-40 C to 1372 C: 1 % of
reading or 1 °C, whichever is
greater
(-40 F to 2501 F: 1 % of
reading or 2 °F, whichever is
greater)