Brochure

The 707Ex is an intrinsically safe loop calibrator for
use in explosion endangered areas. The Fluke 707Ex
is certified in accordance with the ATEX directive (Ex
II 2 G Eex ia IIC T4) in Zones 1 and 2. The 707Ex is
also certified to operate in accordance with Factory
Mutual N.I. Class 1, Division 2 areas Group A-D.
Large display and simple, quick-click push rotary button for
easy one-handed operation.
Simultaneous mA and % readout for quick, easy,
interpretation of readings.
mA accuracy of 0.015%, superior to other loop calibrators.
1 µA resolution for mA source, simulate and measure.
Push button with 25% steps for fast, easy linearity checks.
0-100% “span check” for fast confirmation of zero and span.
Selectable slow ramp, fast ramp, and step ramp to provide
smooth outputs for valve slewing and loop functional tests.
Internal loop supply, so you can power and read a
transmitter at the same time without carrying a DMM.
Measures V dc to 28 V.
Single 9 V battery that is easily changed.
0-20 mA or 4-20 mA default start up modes.
Innovative output adjustment dial with 1 µA and 100 µA
resolution.
HART
®
compatible resistance is connected in series with the
loop supply for compatibility with HART communicators.
Function Range Resolution Accuracy
Voltage measure 0 to 28 V 0.001 V ± (0.015% Rdg + 2 counts)
mA measure 0 to 24 mA 0.001 mA ± (0.015% Rdg + 2 LSD)
mA source
1
0 to 24 mA 0.001 mA ± (0.015% Rdg + 2 LSD)
mA simulate
2
0 to 24 mA 0.001 mA ± (0.015% Rdg + 2 LSD)
Loop supply 24 V dc n/a 24 V ± 1 V dc, no load
Temperature Coefficient, -10 to 18 ºC, 28 to 55 ºC, ± .005% of range per ºC;
1
Max load, 700 Ohms at 20 mA
2
Max applied voltage for simulation, 28V
Specifications
Summary Specifications (18º to 28ºC, one year)
®
The fast, one-handed tool for loop
checks in Ex zones
Fluke 707Ex Intrinsically Safe
Loop Calibrator

Summary of content (2 pages)