HART Compatibility
When in source mode the Calibrator has > 250 Ohms of
series resistance and is compatible with HART devices
without needing an additional series resistor.
Measuring dc Volts
W Caution
To prevent damage to the unit under test,
ensure that the Calibrator is in the correct
mode before connecting the test leads.
To measure dc Volts:
1. Press A to step to MEASURE mode.
2. MEASURE V is displayed.
3. Touch test lead probes across the load or power
source. Connect the COM probe first.