User Manual

707Ex
Users Manual
14
Accuracy Specifications
Accuracy is specified for 1 year after calibration at
operating temperatures of 18 °C to 28 °C and is given as:
± ( [ % of reading ] + [ counts ] )
MEASURE V dc
Range: + 28 V max
Resolution: 1 mV
Input Impedance: 1 M
Accuracy: ± (0.015 % of reading + 2 counts)
MEASURE mA dc
Range: 20 mA (24 mA max)
Resolution: 1 µA
Accuracy: ± (0.015 % of reading + 2 counts)
SOURCE / SIMULATE mA dc
Range: 0 mA to 20 mA (24 mA max)
Resolution: 1 µA
Accuracy: ± (0.015 % of reading + 2 counts)
Source mode:
Compliance: To 700 at 20 mA
Simulate mode:
External loop voltage requirement: 24 V nominal, 28 V
maximum, 12 V minimum
Loop Power
Maximum Load: 700
Percent display
25 % to 125 %
Input / Output Protection
Fused; not replaceable
General Specifications
Maximum voltage between any terminal and earth
ground or between any two terminals:
28 V
Storage temperature:
30 °C to 60 °C
Operating temperature: