User Manual

mA Calibrator
General Specifications
15
Operating altitude:
3000 meters maximum
Temperature coefficient:
± 0.005 % of range per °C for temperatures of 10 °C to
18 °C and 28 °C to 50 °C
Relative humidity:
95 % up to 30 °C;
75 % up to 40 °C
45 % up to 50 °C
Vibration:
Random 2 g, 5 to 500 Hz
Shock:
1 meter drop test
Safety Compliance:
Complies with IEC 61010-1-95 CAT I, 28 V; CSA C22.2
No. 1010-92 NRTL; ANSI/ISA S82.02.01-1994;
( Directive 94/9/EG and p NEC 500: Uo = 27.6 V,
Io = 96.13 mA, Co = 76 nF, Lo = 2.5 mH, Ui = 30 V,
Ii = 24 mA, Ci = 10 nF, Li = 0 mH
CE:
Complies with EN61010-1 and EN61326
Power requirements:
Single 9 V battery
(See Approved Batteries table in Replaceable Parts
section, page 13)
Battery life (typical):
SOURCE mode: 18 hours; 12 mA into 500 ;
MEASURE / SIMULATE mode: 50 hours
Size:
69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H)
[ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ]
With red holster and Flex-Stand:
76.20 mm (W) x 158.75 mm (L) x 54.61mm (H)
[ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ]
Weight:
0.28 kg (0.62 lb)
With red holster and Stand: 0.42 kg (0.93 lb)