719 Series Pressure Calibrator with Electric Pump Users Manual August 2008 © 2008 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years (one year for pump assembly) from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
Table of Contents Title Page Introduction .................................................................................................................... Safety Information .......................................................................................................... Getting Acquainted with the Calibrator ........................................................................... HART Resistor................................................................................................
719 Series Users Manual Sourcing 4 to 20 mA ...................................................................................................... Simulating a 4 to 20mA Transmitter............................................................................... Percent Error Setup ....................................................................................................... Maintenance ..................................................................................................................
List of Tables Table 1. 2. 3. 4. 5. 6. 7. Title Page Symbols................................................................................................................................. Front Panel Features............................................................................................................. Pushbutton Functions............................................................................................................ Pump Features..............................................
719 Series Users Manual iv
List of Figures Figure 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. Title Page Connection Technique .......................................................................................................... Front Panel Features............................................................................................................. Pump Features...................................................................................................................... Internal Pressure Sensor with Internal Pump .....
719 Series Users Manual vi
Introduction The Calibrator features: The 719 30G and 100G Series Pressure Calibrators (the Calibrator) perform the following functions: • Precision air bleed for fine control • Fine vernier adjustment for the pump • Calibrate P/I (pressure to current) transmitters • A serviceable electric pump • Calibrate I/P (current to pressure) devices • Pressure limit settings • Identify the set, reset and deadband of of pressure switches • HART resistor mode • Measure pressure via a 1/8-in.
719 Series Users Manual The Calibrator makes 5-digit pressure measurements in the units shown below: • Psi • inH2O at 4 °C • inH2O at 20 °C • kPa • cmH2O at 4 °C • cmH2O at 20 °C • bar • mbar • kg/cm2 • inHg • mmHg For pressure modules, full-scale readings for all pressure ranges can be made in: • Psi • kPa • inHg To avoid display overflow, full scale readings are limited to 1000 psi in cmH2O, mbar, and mmHg units, and 3000 psi in inH2O units.
Pressure Calibrator with Electric Pump Introduction • • • • • • • • • • • Make sure the battery door is closed and latched before operating the Calibrator. Do not operate the Calibrator if it is damaged. Do not operate the Calibrator around explosive gas, vapor, or dust. When using probes, keep fingers behind the finger guards on the probes. Use only two 9 V batteries, properly installed in the Calibrator case, to power the Calibrator. Follow all equipment safety procedures.
719 Series Users Manual Table 1. Symbols Symbol J fgx001f.eps Figure 1. Connection Technique 4 Earth ground M Battery W Caution: Important information. Refer to instruction sheet X T Hold in fixed position Meaning Hazardous voltage. Risk of electric shock. Double insulated ) Conforms to relevant Canadian Standards Association directives. P Conforms to European Union requirements f Pressure ~ Do not dispose of this product as unsorted municipal waste.
Pressure Calibrator with Electric Pump Getting Acquainted with the Calibrator Getting Acquainted with the Calibrator The Calibrator displays pressure and current measurements simultaneously. See Table 2 and Figure 2 for front panel features. The upper display shows the applied pressure or vacuum (shown as a negative value). Press ENTER then UNITS to select a different unit. When the power is cycled off and on, the Calibrator retains the unit last selected.
719 Series Users Manual Table 3. Pushbutton Functions Pushbutton Z UNITS Description Press to zero the pressure display. Vent pressure to atmosphere before pressing. With an Absolute Pressure Module, see special instructions below. Press C then UNITS to change pressure units. Continue to press UNITS for next selection or use U to move back or V to move forward. Press ENTER when finished or await timeout. All units are available when the pressure sensor input is used.
Pressure Calibrator with Electric Pump Getting Acquainted with the Calibrator Table 4. Pump Features Item A Description MINAX M mA UT OUTP mA Internal Pump- Press P to activate internal electric pump and source pressure/vacuum. L. ST AL ADJU ITCH SWTE ST P PUM 25% Step ITS LIM 30VAX M B Pressure Vacuum Switch- Rotate forward (clockwise) for pressure, backward (counterclockwise) for vacuum.
719 Series Users Manual HART Resistor The Calibrator has a selectable 250 Ω HART resistor to facilitate use with HART communication devices. Use a HART communicator when measuring mA with loop power or sourcing mA. The HART resistor defaults to OFF. Switch Test To perform a switch test: To turn on the HART resistor: 1. With the Calibrator OFF, press O. 2. When HArt is displayed, press Vor U to turn on/off. 1. Power Saver The Calibrator automatically turns off after 30 minutes of inactivity.
Pressure Calibrator with Electric Pump Zeroing with Absolute Pressure Modules 3. 4. 5. 6. Press S to enter pressure switch test mode. The Calibrator will display CLOSE instead of a mA measurement. Apply pressure with the pump slowly until the switch opens. Note In the switch test mode, the display update rate is increased to help capture changing pressure inputs. Even with the enhanced sample rate, pressuring the device under test should be done slowly to ensure accurate readings.
719 Series Users Manual Set Maximum Pressure Limit Using the Internal Pump To set a maximum pressure limit for the internal electric pump: The internal pump can provide the rated pressure for the Calibrators. 1. Press C then LIMITS and the limit setting is displayed. 2. Use Z to increase or Y the limit setting. The preferred use for the internal pump is shown in Figure 4, where the Calibrator displays pressure measured with the internal sensor and provided by the internal pump. 3.
Pressure Calibrator with Electric Pump Using the Internal Pump To use the Calibrator’s internal pump, refer to Figure 3 and perform the following steps: 1. Depressurize and drain the line before connecting the Calibrator. 2. Connect the pressure transmitter to the Calibrator internal sensor as shown in Figure 4 (for internal pressure sensor measurements) or Figure 5 (for pressure module measurements). 3. 4. Note To avoid leaks, use Teflon tape or similar sealant on all pressure connections.
719 Series Users Manual 719 RE SSU TOR PRE LIBRA CA Black HOLD P LOOWER PO Red ER O ENT ZER S UNIT MINAX M mA UT OUTP mA L. ST AL ADJU ITCH SWTE ST HOLD P PUM 25% Step mA ITS LIM 30VAX M COMTE ULA /SIM RCE 24 mA 0- S I G N A L SOU + – T E S T fgx002f.eps Figure 4.
Pressure Calibrator with Electric Pump Using the Internal Pump S I G N A L Pressure Module + – T E S T 719 RE SSU TOR PRE LIBRA CA HOLD P LOOWER PO ER O ENT ZER S UNIT MINAX M mA UT OUTP mA L. ST AL ADJU HOLD P PUM 25% Step f mA ITS LIM V 30VA 30 MAXX M COMTE ULA /SIM RCE 24 mA 0- SOU -7 Red Black fgx010f.eps Figure 5.
719 Series Users Manual Table 5.
Pressure Calibrator with Electric Pump Pump Valve Assembly Cleaning Instructions Pump Valve Assembly Cleaning Instructions 9. 1. Using a small screwdriver, remove the two valve retention caps located in the oval shaped opening on the underside of the Calibrator. 10. Seal the output of the Calibrator and pump up the unit to at least 50 % its rated pressure. 2. After the caps have been removed, gently remove the spring and o-ring assembly. 3.
719 Series Users Manual S I G N A L Pressure Module + – 719 RE SSU TOR PRE LIBRA CA HOLD T E S T P LOOWER PO ER O ENT ZER S UNIT MINAX M mA L. ST AL ADJU ITCH SWTE ST P PUM 25% Step f mA ITS LIM 30VAX M COM TE ULA /SIM RCE 24 mA 0- SOU Red Black fgx006f.eps Figure 6.
Pressure Calibrator with Electric Pump External Fluke Pressure Module Compatibility External Fluke Pressure Module Compatibility If inappropriate units are selected, the output of Fluke 700P Pressure Modules can cause the Calibrator display to overflow (OL), or displays values that are too low to be read. Refer to Table 6 for appropriate unit and range compatibility. Table 6.
719 Series Users Manual Supplying Loop Power mA Modes The Calibrator can supply loop power at 24 V dc to a current transmitter that is disconnected from the system. Use the following procedure: Different mA functions can be accessed by repeatedly pressing A: 1. With power off, hold down Z while pressing n. Loop Power appears in the display. 2.
Pressure Calibrator with Electric Pump Sourcing 4 to 20 mA Sourcing 4 to 20 mA Simulating a 4 to 20mA Transmitter To select current-sourcing mode, use the following procedure: Simulate is a mode of operation in which the calibrator is connected into a loop in place of a transmitter and supplies a known, settable test current. 1. Press A until Source is displayed. 2. Connect leads as shown in Figure 8. 3. Enter the desired current by pressing Z or Y. 1.
719 Series Users Manual 719 PRESSURE CALIBRATOR Red + Loop Power Supply +24 V Black Readout or Controller – fgx011.eps Figure 9.
Pressure Calibrator with Electric Pump Percent Error Setup Percent Error Setup In Case of Difficulty 1. Press and hold A. After 3 seconds the set icon and 0% appears on the lower display. • 2. Use Vand U to adjust the % point for the Percent Error calculation, then press ENTER to confirm selection. Check the battery, test leads, pressure module, and pressure tubing. Follow replacement and connection instructions properly. • Review this manual to make sure the Calibrator is used correctly. 3.
719 Series Users Manual Replacing the Batteries When the battery symbol (B) appears, replace the two 9 V alkaline batteries. Refer to Figure 10. XW Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator (B ) appears. Remove test leads before changing the battery. Calibration Fluke recommends calibrating the Calibrator once a year to ensure that it performs according to its specifications. wh008f.
Pressure Calibrator with Electric Pump Parts and Accessories Parts and Accessories Refer to Table 7 and Figure 11. Table 7. Replacement Parts Item Description Part/ Mod. No.
719 Series Users Manual Item Description Part/ Mod. No.
Pressure Calibrator with Electric Pump Parts and Accessories MP85 MP1 Holster AC72 Alligator Clips MP2 S1 MP5 TM1 TL75 Test Lead Set MP14 MP11,12,13 H7,8 MP7 MP3,4 MP8 MP6 MP86 BT1 BT2 MP92 MP9 H5,6 H2,3,4 CD-ROM MP89,90 fgx004f.eps Figure 11.
719 Series Users Manual Specifications Specifications are based on a one year calibration cycle and apply for ambient temperature from +18 °C to +28 °C unless stated otherwise. Counts are the number of increments or decrements of the least significant digit. Pressure Sensor Input Model 30G 100G Range DC mA Measure and Source Accuracy -12 to 36.0 psi ± 0.025 % of Range (6 month calibration) -12 to 120.0 psi ± 0.
Pressure Calibrator with Electric Pump Contacting Fluke General Specifications Maximum voltage applied between either mA terminal and earth ground or between the mA terminals: 30 V Storage temperature: -30 °C to 60 °C Operating temperature: -10 °C to 55 °C Operating altitude: 3000 m maximum Relative humidity: 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to 50 °C, and 35 % up to 55 °C Vibration: Random 2 g, 5 Hz to 500 Hz per MIL-PRF28800F Class 2 Calibration Cycle: 6 months to 2 years based on desired accur
719 Series Users Manual 28
Manual Supplement Manual Title: Print Date: Revision/Date: 719 Users August 2008 Supplement Issue: Issue Date: Page Count: 2 3/09 2 This supplement contains information necessary to ensure the accuracy of the above manual. This manual is distributed as an electronic manual on the following CD-ROM: CD Title: CD Rev. & Date: CD PN: 719 8/2008 3316449 © 2008-2009 Fluke Corporation. All rights reserved. Printed in the U.S.A.
719 Users Manual Supplement Change #1, 48042 and 49682 On page 26, replace the Pressure Sensor Input and the DC mA Measure and Source tables with the following: Pressure Sensor Input Model Range Resolution Accuracy [1] Max Non-destructive Pressure 30G -12 to 30.0 psi 0.001 ±0.025 % of Range (6 month calibration) 60 psi 100G -12 to 100.0 psi 0.01 ±0.035 % of Range (1 year calibration) 200 psi Temperature coefficient: 0.
Manual Supplement 719 Users DC mA Measure and Source Range 24 mA Accuracy ±(% of Reading + Counts) [1] Resolution 0.001 mA 0.015 + 2 Maximum load on mA source is 1000 Ω. With HART resistor on, maximum load is 750 Ω. Fuseless overload protection Temperature coefficient: 0.005 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C [1] In an RF field of 1V/m to 3V/m and between the frequencies of 150 MHz to 250 MHz, add 75 counts to specified accuracy.